Научная статья на тему 'C-MEMS Derived Glassy Carbon Glucose Sensor'

C-MEMS Derived Glassy Carbon Glucose Sensor Текст научной статьи по специальности «Нанотехнологии»

CC BY
161
15
i Надоели баннеры? Вы всегда можете отключить рекламу.
Ключевые слова
C-MEMS / glassy carbon / biosensor / chitosan / glucose oxidase / impedance spectroscopy / electrochemical sensing / bio-functionalization

Аннотация научной статьи по нанотехнологиям, автор научной работы — Naresh Mandal, Bidhan Pramanick

Diabetes is one of the most severe health diseases in the world. More than 150 million people worldwide suffer from diabetes caused by an abnormal glucose concentration in their blood and tissues. The analysis of glucose levels in the body is significant, primarily due to diabetes mellitus. Therefore, accurately detecting blood glucose is crucial for diagnosing, treating, and managing diabetes. Diabetic patients can self-manage their condition by monitoring their glucose levels. The electrochemical glucose biosensor has several advantages, including recognizing glucose specifically, low price, ease of sensor manufacture, correctness, portability, and easy operation. We have developed glassy carbon electrodes (GCE) with conventional carbon-micro-electromechanical (C-MEMS) procedures for glucose sensing. SU-8 photoresist was used as the carbon precursor. The fabricated C-MEMS-derived GCE surface has been functionalized with chitosan and glucose oxidase. Glucose oxidase is a well-known enzyme for oxidizing glucose into gluconic acid and H2O2. This reaction kinetics has been recorded with increasing glucose concentration using electrochemical analysis. In summary, we have presented an electrochemical glucose biosensor in a single-step immobilization protocol of glucose oxidase on the glassy carbon electrode (GCE) surface and obtained satisfactory responses for 1 mM to 10 mM glucose solutions. © 2023 Journal of Biomedical Photonics & Engineering.

i Надоели баннеры? Вы всегда можете отключить рекламу.
iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.
i Надоели баннеры? Вы всегда можете отключить рекламу.

Текст научной работы на тему «C-MEMS Derived Glassy Carbon Glucose Sensor»

C-MEMS Derived Glassy Carbon Glucose Sensor

Naresh Mandal1 and Bidhan Pramanick1,2*

1 School of Electrical Sciences, Indian Institute of Technology Goa, Ponda 403401, India

2 Centre of Excellence in Particulates Colloids and Interfaces, Indian Institute of Technology Goa, Ponda 403401, India

*e-mail: bidhan@iitgoa.ac.in

Abstract. Diabetes is one of the most severe health diseases in the world. More than 150 million people worldwide suffer from diabetes caused by an abnormal glucose concentration in their blood and tissues. The analysis of glucose levels in the body is significant, primarily due to diabetes mellitus. Therefore, accurately detecting blood glucose is crucial for diagnosing, treating, and managing diabetes. Diabetic patients can self-manage their condition by monitoring their glucose levels. The electrochemical glucose biosensor has several advantages, including recognizing glucose specifically, low price, ease of sensor manufacture, correctness, portability, and easy operation. We have developed glassy carbon electrodes (GCE) with conventional carbon-micro-electromechanical (C-MEMS) procedures for glucose sensing. SU-8 photoresist was used as the carbon precursor. The fabricated C-MEMS-derived GCE surface has been functionalized with chitosan and glucose oxidase. Glucose oxidase is a well-known enzyme for oxidizing glucose into gluconic acid and H2O2. This reaction kinetics has been recorded with increasing glucose concentration using electrochemical analysis. In summary, we have presented an electrochemical glucose biosensor in a singlestep immobilization protocol of glucose oxidase on the glassy carbon electrode (GCE) surface and obtained satisfactory responses for 1 mM to 10 mM glucose solutions. © 2023 Journal of Biomedical Photonics & Engineering.

Keywords: C-MEMS; glassy carbon; biosensor; chitosan, glucose oxidase, impedance spectroscopy; electrochemical sensing; bio-functionalization.

Paper #6476 received 13 Feb 2023; revised manuscript received 20 May 2023; accepted for publication 21 May 2023; published online 1 Aug 2023. doi: 10.18287/JBPE23.09.030305.

1 Introduction

In the 21st century, diabetes has become a significant disease around the globe, with more than 439 million cases of Diabetes Mellitus (DM) estimated in 2030 [1]. According to WHO, 422 million people worldwide suffer from diabetes mellitus, which is predicted to be the seventh deadliest disease by 2030 [2]. Long-term suffering from DM causes cardiovascular diseases, kidney failure, blindness etc. The only way to cure DM is to control glycaemia levels. To control the glucose level in the blood, a diabetic patient needs to monitor the blood glucose level regularly. Therefore, it is a need for the hour to develop a novel, quick-response, low-cost biosensor for glucose sensing. Nano glucose sensors are usually made using an oxidoreductase to oxidize the glucose on the electrode surface. In general, glucose

oxidase (GOx) is used for this purpose because it has excellent glucose selectivity, a high degree of stability, and a high activity level across a wide pH range [3]. According to this research, when GOx is used directly with bare electrode surfaces, it changes its structure, biological activity, and function [4]. The conventional finger-pricking method of measuring blood glucose levels causes pain, infection, and tissue damage [5]. The alternative way of finger pricking is without using blood samples in a non-invasive way [6], using bio fluids like tears, sweat, interstitial fluid (ISF), and saliva [7]. The non-invasive measurement techniques include the electromagnetic wave (EM) method, optical measuring and electrochemical sensing. M. Gourzi et al. developed a technique for measuring blood glucose from the skin using EM waves. However, this method is unreliable because measured glucose levels depend on temperature

This paper was presented at the International Conference on Nanoscience and Photonics for Medical Applications - ICNPMA, Manipal, India, December 28-30, 2022.

and cholesterol and ionize the biomolecules [8]. The mid-infrared spectroscopy is better than near-infrared spectroscopy [9]; in both methods, the pressure, temperature, and humidity severely affect the measured glucose value from the skin [10]. The temperature changes influence the measured value of glucose from the skin by optical coherence tomography (OCT) [11]. The specificity of the optical polarimetry technique of glucose measurement from the eye is very low because of eye movement, motion artifact and the optically active composition of aqueous humor [12].

A great deal of interest has been shown in enzyme immobilization technology concerning designing, fabricating, and applying biosensors. Various immobilization techniques are available, such as physical methods (absorption or entrapment) and chemical methods (for example, covalent bonds). Detecting trace components of biological, clinical, industrial and environmental systems can be carried out relatively simply, sensitively, and selectively using electrochemistry. When preparing biosensors, immobilization on the solid matrix is vital to maintaining enzyme function. It is essential that the structure and thickness of the materials used to immobilize enzymes do not reduce their functionality severely.

In the case of chitosan (CS), this polysaccharide biopolymer is formed by the partial deacetylation of chitin. The biocompatibility of chitosan makes it a desirable structural material for enzymatic immobilization and enhances biosensors' performance by retaining enzyme activity. Using chitosan for immobilizing enzymes and cells has been demonstrated as a functional, structural material. Several papers have reported using chitosan-modified biosensors, for example, the modified glucose biosensor built using a sol-gel/chitosan membrane [13, 14]. Adhesion and non-toxicity are good characteristics of this material. Additionally, its structure contains amino and hydroxyl groups, allowing easy chemical modification [15].

The electrochemical method of sensing glucose has immensely made the process easy due to its quick response, less sample use and reduced cost of testing. It is possible due to the advancement of low-cost fabrication techniques of sensors [16]. There are two methods of electrochemically sensing glucose from biofluids such as saliva, sweat, and tears: enzymatic and non-enzymatic. Clark and Lyons first evolved the concept of enzymatic bio-modified electrode sensor [17]. There have been numerous attempts to improve electron transfer from glucose oxidase (GOx) to electrode surfaces. As a result, there has been a growing interest in using semiconductors, carbon materials, such as carbon nanotubes, graphene, glassy carbon (GC) etc., and conductive polymers for achieving this goal. A fabricated platinum electrode was bio-modified in the layer-by-layer structure of single-walled carbon nanotube (SWNT), chitosan, and gold nanoparticles to measure glucose from saliva, obtaining a low detection limit of 0.1 mg/dl [18]. H. Yao et al. developed a contact lens sensor system to measure enzymatically tear glucose

in vitro [19]. A study was carried out to measure glucose from ISF in tattoo based, noninvasive way [20]. Z. Temogin implemented a sensor by modifying the surface of glassy carbon electrodes (GCE) with a blend nanofiber of poly (vinyl alcohol) and poly (ethylenemine), immobilized glucose oxidase (GOx). They obtained linear responses for blood glucose range, long storage stability, and high anti-interference for ascorbic and uric acid [21]. A GCE functionalized with high catalytic MnO2 nanowires and coated with GOx showed high sensitivity [22]. Followed by modified GCE with MnO2NPs and carbon nanofiber (CNF) nanocomposites, which showed higher performance in terms of sensitivity, selectivity, and response time [23]. The electron transfer rate was enhanced in a GCE modified by Co3O4 nanoparticles, multiwalled carbon nanotubes (MWCNTs) and immobilized GOx [24]. It was also experimentally verified that Ni/CoO-loaded carbon nanofiber deposited GCE showed a good catalytic effect on glucose [25]. To enhance the performance of glucose sensors, GCE was modified by polypyrrole, nafion functionalized Multiwalled carbon nanotubes (MWCNTs) nanocomposite, chitosan and glucose oxidase showed excellent catalytic activity [26]. GOx, encapsulated in liposome microreactors, was coated on a chitosan layer over GCE showing excellent repeatability and stability [27]. The non-covalently functionalization of MWCNTs and GOx was also deposited on GCE to detect glucose in milk [28]. The poly(p-phenylenediamine) (PPD) and Fe3O4 nanoparticles (Fe3O4 NPs) nanocomposites were developed as enzyme immobilizers on GCE. The modified electrode showed good affinity to glucose in clinical samples without interferences [29]. The GOx was immobilized on single-layer graphene doped with AuNPs over fabricated GCE. The sensor showed ultrasensitive for low concentrations of glucose [30]. The 2D-RGO (Reduced graphene oxide) nanocomposites modified GCE, developed by G. Bharath et al., obtained higher electron transfer rate and sensitivity [31]. It has been demonstrated that palladium nanoparticle/chitosan-grafted graphene nanocomposites can be used as glucose biosensors [32]. To detect dopamine selectively, graphene/chitosan electrodes were developed [33]. In direct electrochemistry and glucose sensing, a glucose oxidase-graphene-chitosan electrode was developed [34].

Sol-gel techniques have been used to construct many of these very successful biosensors. However, according to our knowledge, the method for producing biosensors also has some disadvantages. For example, through the porous nature of the thin sol-gel matrix, the enzyme is slowly leached from the electrode surface. There are several issues with the biosensor, such as its slow response time, the procedures being more complicated, and a fragility that makes it difficult to use. It is important to note that during the sol-gel process, certain harsh conditions are exposed to the enzyme, namely partial denaturation by the catalyst (ether basic or acidic) or the solvent (usually methanol).

Fig. 1 General scheme of bio-modified glassy carbon electrode (GCE) sensor.

In this paper, a novel biosensor is presented to overcome these problems. We have developed a simple, cost-effective glucose biosensor based on the entrapment of glucose oxidase directly in chitosan on a GC electrode. Compared to the sol-gel/chitosan-modified Glucose Oxidase biosensor, this low-cost design simplifies the fabrication process to ensure a stable and sensitive biosensor. A significant advantage of such devices is that they offer faster, more straightforward methods of obtaining sequence-specific information than the traditional sol-gel techniques. An optimization study was conducted for the biosensor, and its analytical performance was evaluated. A schematic representation of the steps involved in the modification process of the glassy carbon electrode (GCE) for the glucose electrochemical sensing platform is shown in Fig. 1.

2 Materials and Methods

2.1 Reagents and Solutions Preparation

For bio-modifying GCE chitosan from shrimp shells, > 75% (deacetylated) Glucose Oxidase from Aspergillus niger with an activity of 100000-250000 units/g was purchased from Sigma Aldrich. Phosphate buffer saline (PBS) contains NaCl, KCl, Sodium dihydrogen phosphate (Na2HPO4), and Potassium dihydrogen phosphate (KH2PO4), which were purchased from Sigma Aldrich. Standard glucose solution is prepared from D-(+)-Glucose > 99.5% (GC) was purchased from Sigma Aldrich. Glacial acetic acid was purchased from Sigma Aldrich to prepare chitosan solution of chitosan in 0.1 M acetic acid. Milli-Q 0.5 water with a resistivity of 18 M^ cm was taken from in house plant of IIT Goa for making PBS solutions and washing the electrodes. Chitosan solution has been prepared by dissolving (2 mg/mL) chitosan sample in 0.6 wt% glacial acetic acid at room temperature using magnetic stirring at 1000 rpm

for 1 h. A glucose oxidase (GOx) concentration of 35 mg/mL has been prepared in 0.25 M PBS.

2.2 Fabrication of Electrodes

The conventional method of photolithography and pyrolysis was used to fabricate GCE (Fig. 2) on a 4-inch silicon wafer with a 1 ^m thermally-oxidized SiO2 layer (NOEL Technologies, Campbell, CA, USA). SU-8 negative photoresist was spin-coated at 3000 rpm for the 30 s and then soft-baked on a hotplate at 95 °C for 3 min to get an even distribution of the layer on the substrate and evaporate the solvent. UV radiation was exposed on photoresist through a thin film photomask to a total energy of 120 mJ/cm2 and baked again for 3.5 min at 95 °C to crosslink the exposed patterns. Then the samples were dipped in SU-8 PR developer solution for 2 min to dissolve the unexposed region [35]. The SU-8 structured pattern was pyrolyzed in an inert furnace to convert the SU-8 to glassy carbon (GC) structure. Initially, the SU-8 pattern was heated to 300 °C at a determined temperature ramp rate of 5 °C/min and kept at that temperature for 60 min for thermal stabilization. Then the temperature was increased to 900 °C at a rate of 5 °C/min, maintained the temperature for 1 h and cooled down the furnace at a rate of 10 °C/min. Thus, the SU-8 sensor structure was pyrolyzed to GC structure [36].

2.3 Electrode Surface Modification

The chitosan solution 0.6 wt% in 0.1 M acetic acid was prepared by magnetic stirring at 1000 rpm for 1 h. The GCE electrode was cleaned in Milli-Q water, dried using N2 gas flow, and coated with chitosan solution by pipetting. Then it was kept at room temperature for 1 h till drying. Afterwards, the chitosan-coated GCE was pipetted with 35 mg/ml GOx solution in 0.25 M PBS and preserved at 4 °C temperature for 2 h before use (Fig. 3).

Fig. 2 Fabrication steps of GC electrode.

Fig. 3 Bio-modifying of GC electrode.

The bio-modified electrode was cleaned in PBS solution to remove unbound glucose oxidase enzyme before electrochemical testing. All electrochemical experiments were carried out using a potentiostat of CHI 760E electrochemical workstations. A three-electrode system of bio-modified GCE as the working electrode, Ag/AgCl as the reference electrode and Platinum wire as the counter electrode was used for electrochemical characterization.

2.4 Structural Characterization

The surface of the patterned SU-8 structure was shrunk due to pyrolysis. Scanning Electron Microscopy (SEM) (EVO, Zeiss, Germany) was used to characterize the surface morphology and to measure the shrinkage of the carbonized electrodes. The crystallinity of the fabricated carbon electrode was examined with Raman spectroscopy (inVia Qontor, Renishaw plc, UK) using a 514 nm Ar laser source. This work has been reported in the previously reported work [36].

2.5 FTIR Characterization

FTIR characterization has been performed using a SHIMADZU (IRAFFINITY 1S WL model) with a resolution of 4 cm-1 in the 400-4000 cm-1 region. This instrument is equipped with Specac attenuated total reflection (ATR) fitted with diamond crystal, scanned (transmittance vs wave numbers mode) at room temperature 25 °C and humidity. Fourteen hundred scans

have been taken for accuracy, and we reported the average value.

2.6 Electrochemical Characterization

Electrochemical impedance spectroscopy (EIS) is an efficient electrochemical measurement technique to analyse the behaviour of the surface-modified electrodes for a particular analyte [1]. The EIS data is usually fitted with an equivalent electrical circuit model to get the electrical fingerprint of the electrochemical behaviour of the sample. EIS was carried out for glucose concentrations 1 mM to 10 mM in 0.25 M PBS, applied ac signal perturbation 5 mV and frequency range from 100 Hz to 1 MHz. Cyclic voltammetry (CV) was carried out to observe the redox peak of bio-modified GCE for glucose concentrations 1 mM, 3 mM, 5 mM, and 8 mM for an applied potential range of -0.1 V to 0.6 V vs Ag/AgCl electrode at a scan rate of 100 mV/sec. We recorded Chronoamperometry reading for 1 mM, 3 mM, 5 mM, 8 mM, and 10 mM glucose concentrations to quantify the glucose concentration. All the experiments were carried out for glucose solutions in 0.25 M Phosphate buffer saline (PBS) of pH 7.4 and room temperature 25 °C. All the readings were taken in triplets for each technique and the average has been plotted.

3 Results and Discussion

3.1 Morphologies of GCE, Chitosan, and Chitosan/ Glucose Oxidase / GCE

Micrographs of chitosan and chitosan/glucose oxidase have been characterized using SEM images, as shown in Fig. 4. Noticeably, Fig. 4(b) shows that the membrane of chitosan consists of a homogeneous arrangement, and its surface is flat and symmetrical.

Fig. 4 Scanning electron microscopy (SEM) photomicrograph (images captured at the magnification of 1.63 K X) showing the morphology of (a) glassy carbon electrode (GCE), (b) chitosan/GCE, and (c) chitosan/ glucose oxidase /GCE.

A bump structure is visible on the chitosan/glucose oxidase/GCE compared to the chitosan/GCE (Fig. 4b). The bumps are uniform in size and shape, suggesting glucose oxidase is immobilized in the chitosan/GCE surface (Fig. 4c).

■ Bare Glassy Carbon Electrode (GCE) (a) GCE+Chitosan (b)

■ GCE+Chitosan+Glucose oxidase (c)

3.2 Characterization of the Sensor Electrode by FTIR

First, 2 mg/mL chitosan solution was pipetted out on the GCE and kept at room temperature for 1 h. After the treatment, FTIR measurement was carried out to estimate the binding of GC with chitosan. After thoroughly drying the surface of the active area of GCE, 10 ^L of glucose oxidase solution was pipetted out and allowed to dry for 2 h at room temperature. Again FTIR measurement has been recorded. For each of the electrochemical processes, FTIR plot of transmittance vs wave numbers have been recorded, which is depicted in Fig. 5(a) glassy carbon electrode (GCE), (b) GCE/chitosan, and (c) GCE/chitosan/ glucose oxidase.

GCE, the bare electrode, does not show many FTIR features that suggest the glassy carbon nature of the electrode (Fig. 5a). After adding chitosan (2 mg/mL) to the GCE, a broad peak appears around 3235 cm-1, corresponding to O-H and N-H stretching frequency. The broadness of the peak is due to substantial hydrogen bonding. The peaks around 2915 and 2872 cm-1 are due to C-H stretching frequency. Carbonyl stretching frequencies around 1631 and 1549 cm-1 are characteristics of amide bonds of chitosan (N-acetyl groups). The most substantial peak at 1061-1023 cm-1 is attributed to C-O stretching frequencies of polysaccharides (Fig. 5b). On adding glucose oxidase to the chitosan-modified GCE, the peak intensities are reduced substantially, suggesting surface modification. A broad peak around O-H and N-H stretching frequency regions (~ 3250 cm-1) was observed. The amide peaks around 1680-1643 cm-1 and 1543-1502 cm-1 were also observed. The prominent peak at 1067 due to C-O stretching was observed for glucose oxidase-modified electrodes (Fig. 5c). Comparing the FTIR of surface-modified electrodes, it is evident that the surface was modified initially by chitosan and subsequently with glucose oxidase. No new peaks were observed as chitosan, and glucose oxidase interaction is primarily hydrogen bonding.

Q>

o

E

ra £

E

c

ro

--- ---(b)

3235^t ^872 / !o23

2915 1061 _________

t / Tl5 ^

3250 1643 15'43&

1067

3600

3000

2400

1300

Wevenumber (cm"^)

1200

600

Fig. 5 FTIR spectrum of (a) glassy carbon electrode (GCE), (b) GCE/chitosan, and (c) GCE/chitosan/glucose oxidase.

3.3 Electrochemical Characterization Results

Electrochemical characterization of GCE has been carried out using 0.25 M Phosphate buffer saline (PBS) of pH 7.4 and room temperature 25 °C. A conventional three-electrode system (CHI Electrochemical work analyzer (CH 760E)) has been employed for all Electrochemical measurements. In this system, we have used the carbon electrode as the working electrode, a platinum wire as the counter electrode and an Ag/AgCl as a reference electrode. The Eqs. below show that glucose oxidase reacts with glucose when oxygen is used as an electron acceptor.

Enzyme(oxidized)+ ft-d-glucose^ Enzyme(reduces) +d-gluconic acid,

Enzyme(reduced)+O2^ Enzyme(oxidized) +H2O2.

(1)

(2)

In the presence of an applied potential of +0.5 V, the following reaction occurs when H2O2 is brought closer to the surface of the anode.

Fig. 6 EIS response of biomodified GCE for 1 mM-10 mM glucose concentrations, (A) zoom view of Nyquist plot for higher frequency ranges 100 Hz-1 MHz and (B) Randle's equivalent circuit model.

HO + 2H ++ • (3)

When the H2O2 redox reaction occurs near the carbon electrode sensor surface, the resulting current increases as glucose concentration increases.

3.3.1 Electrochemical Impedance Spectroscopy (EIS) Measurements

During EIS (Electrochemical Impedance Spectroscopy) measurements, a 3-electrode configuration and a PC interfaced a conventional three-electrode system (CHI Electrochemical work analyzer (CH 760E)) was used with an applied sinusoidal signal of 5 mV amplitude in the frequency range of 100 Hz to 1 MHz. The Nyquist impedance plots have been recorded after the GCE has been exposed to 0.25 M Phosphate buffer saline (PBS) of pH 7.4 and room temperature 25 °C. The Nyquist plots of the EIS data in Fig. 6A showed that with the increase in glucose concentration, the radius of the semi-circle was reduced, indicating the decrease in Rct in Randle' s circuit model Fig 6B.

The current through the electrode consists of faradaic current and double-layer capacitor charging current. Therefore, the electrical equivalent model can be defined as solution resistance (Rs) in series with the parallel combination of double-layer capacitance and in series with charge transfer resistance (Rct) and warbag resistance (Zw), responsible for mass transfer [37]. The warbag resistance (Zw) is associated with diffusion, migration and convocation of charges towards the

electrode [38]. The double layer capacitance (Cdl) is easily formed on GCE due to the porous surface [39]. The value of Cdl depends on the ionic concentration, pH, temperature, type of ions, oxide layers, and roughness of the electrode [40]. The impedance magnitude is decreased with the increase in glucose concentration, as the reactant will be more for higher concentrations. The concentrations measured in this study ranged from 1 mM to 10 mM, indicating that the whole process was controlled by the enzyme's catalytic kinetics. Increasing glucose concentration resulted in more hydrogen peroxide(H2O2) being released from enzyme base reactions and under air saturation conditions; as hydrogen peroxide (H2O2) concentration increases at the GC electrode surface, the oxidation current increases.

3.3.2 Cyclic Voltammetry (CV) Measurements

Cyclic voltammetry (CV) is a versatile electrochemical technique to study the redox kinetics on electrode surfaces with bio analytes. Cyclic voltammetry (CV) measurements of the GCE sensor have been performed in 0.25 M Phosphate buffer saline (PBS) of pH 7.4 at room temperature 25 °C. For Cyclic voltammetry (CV) measurements, a three-electrode system has been used in a similar manner. Using this system, we used the GCE sensor as a working electrode, the platinum wire as a counter electrode, and the Ag/AgCl as a reference electrode. Cyclic voltammetry (CV) was carried out with an electrode potential range of -0.1 V to 0.6V vs Ag/AgCl electrode at a scan rate of 100 mV/sec.

-1.0 -0.8 -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8 Potential (V) vs. Ag/AgCl

Fig. 7 Cyclic voltammogram of biomodified GCE for glucose concentrations 1 mM, 3 mM, 5 mM, 8 mM, and 10 mM.

30 40 50

Time (sec)

Fig. 8 Chronoamperometry response of bio-modified GCE. (A) The current change of different concentrations of glucose solution and the associated fitting curve.

The redox reaction on the top surface of the electrode occurs at a specific potential. It is related to the surface area, rate constant, interface with solution, pH of the solution, and temperature [41]. The peak current of CV responses was increased with the increase in glucose concentrations. The peak current is validated by Randles-Sevick Eqs. (4), peak current is directly proportional to the concentrations [42].

3 11

Ip = (2.69 X 10s)rnAD2Cv2. (4)

Here, C is the concentration, and other symbols have standard meanings. From the CV response, it is obtained that maximum oxidation occurs at 0.5 V. The change of peak current of CV response with the change of concentration is shown in Fig. 7.

(a) (b) (c)

Fig. 9 (a) Variations of chronoamperometry measurements conducted for the glucose concentrations of 1, 3, 5, 8 and 10 mM for successive five days. (b) The measured current variation as a function of storage time for glucose concentration of 1 mM. (c) The reproducibility testing of three GEC sensors.

3.3.3 Chronoamperometery Measurements

The Michaelis-Menten constant is mainly utilized for the biological activity of immobilized glucose oxidase, and the Michaelis-Menten Eq. (5) can be used to calculate it.

I = Imax-KS$, (5)

where I represents the steady-state catalytic current, IMAX represents the maximum current (under saturated substrate conditions), C denotes the glucose concentration, and Ks represents the Michaelis-Menten constant of the system. As shown in Fig. 8, the glucose biosensor shows a typical current-time curve under optimized experimental settings. Chronoamperometery current was recorded for 1 mM-10 mM glucose solution by applying dc voltage 0.5 V. Fig. 8A shows that amperometric current increases with the increase of glucose concentration. The sensitivity obtained by linearizing the stable amperometric current with R2 = 0.8966 curve is 17 ^A/mM. It was observed that when 0.5 V was applied, an amperometric response was well-defined, stable, and fast. There was a direct relationship between glucose concentration and the currents produced in the solution when glucose was added. It took about 10 s for the current value to reach a steady state. This study evaluated 1 mM to 10 mM, which demonstrated that the enzyme's catalytic kinetics governed the whole system.

3.3.4 Stability and Reproducibility Test of the Bio-Modified GCE Sensor

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

Stability is a crucial factor when evaluating the effectiveness of an electrochemical biosensor. The stability of the GCE was assessed using two different methods. First, the GCE was employed to measure glucose levels over five consecutive days, and the resulting chronoamperometric current responses were recorded to determine the stability. The measurements were taken using 1, 3, 5, 8, and 10 mM glucose

concentrations at pH 7.4. Followed by the measures, the GCE was stored in a refrigerator at 4.0 °C. The current values in Fig. 9(a) demonstrate that the GCE exhibited excellent stability in accurately measuring various glucose concentrations throughout the five days.

Ensuring long-term stability in enzyme-based biosensors is a challenging task due to the inherent instability of enzymes. Therefore, the GCE was stored in a refrigerator for 56 days to evaluate its stability further. In the stability test, chronoamperometric current measurements were taken at specific time intervals using glucose solutions with a consistent concentration of 1 mM. Fig. 9(b) illustrates the current values obtained from these measurements over the storage period. It was observed that there was a maximum decrease of 4% and an increase of 3% compared to the initial value. These findings indicate that the GCE exhibits exceptional long-term stability, which can be attributed to the adequate protection of the GOx enzyme through the single-step immobilization process.

The reproducibility of the analytic signal is a crucial requirement for biosensors. Three separate GCE sensors were prepared and utilized to detect glucose in 1 mM glucose solutions to assess its stability. The chronoamperometric current was recorded for each sensor with 1 mM glucose solution. The reproducibility of the different glucose sensors was determined and presented in Fig. 9(c). By calculating the relative standard deviation of glucose concentrations through consecutive measurements, it was found to be 5%. This result indicates that the bio-modified GCE sensors yielded glucose analysis outcomes with relatively low variability, demonstrating a high level of reproducibility.

4 Conclusions

The research on the development of blood glucose measurement in a non-invasive way has received enormous attention in the past two decades as this has become an utmost priority to monitor patients' blood glucose levels regularly. This work presents a thorough study of the available glucose measurement techniques

and their limitations, followed by the development of a glassy carbon glucose sensor. GCE has been fabricated by a conventional photolithography process and carbonization of SU-8 negative photoresist by pyrolysis. The GCE was functionalized by chitosan and glucose oxidase to detect glucose. Glucose oxidase converts glucose into gluconic acid and H2O2, and this reaction kinetics has been measured with an increasing glucose concentration of 1 mM to 10 mM glucose solutions. We have obtained a sensitivity of 17 ^A/mM. We have presented and discussed the electrochemical glucose biosensor in this manuscript, which lies in a single-step immobilization protocol of glucose oxidase on the glassy carbon electrode (GCE) surface.

Acknowledgements

The authors would like to thank the technical staff of MEC and MEMS lab, Dept. of E&ECE, IIT Kharagpur, India, for their support in the fabrication of the carbon electrodes. The authors also would like to thank Dr. Raja Mitra, SCMS, IIT Goa, for his help analyzing FTIR results. The FTIR analysis support from IIT Goa is acknowledged. The authors would like to thank IIT Goa for its financial support.

Disclosures

The authors declare no conflict of interest.

References

1. J. E. Shaw, R. A. Sicree, and P. Z. Zimmet, "Global estimates of the preva lence of diabetes for 2010 and 2030," Diabetes Research and Clinical Practice 87(1), 4-14 (2010).

2. P. López-Soriano, P. Noguera, M. T. Gorris, R. Puchades, Á. Maquieira, E. Marco-Noales, and M. M. López, "Lateral flow immunoassay for on-site detection of Xanthomonas arboricola pv. pruni in symptomatic field samples," PLOS ONE 12(4), e0176201 (2017).

3. S. Ferri, K. Kojima, and K. Sode, "Review of glucose oxidases and glucose dehydrogenases: A Bird's eye view of glucose sensing enzymes," Journal of Diabetes Science and Technology 5(5), 1068-1076 (2011).

4. K.-J. Huang, L. Wang, J. Li, T. Gan, and Y.-M. Liu, "Glassy carbon electrode modified with glucose oxidase-graphene-nano-copper composite film for glucose sensing," Measurement 46(1), 378-383 (2013).

5. J. Chung, H. So, Choi, and T. K. S. Wong, "Recent advances in noninvasive glucose monitoring," Medical Devices: Evidence and Research 5, 45-52 (2012).

6. O. S. Khalil, "Noninvasive glucose measurement technologies: An update from 1999 to the dawn of the New Millennium," Diabetes Technology and Therapeutics 6(5), 660-697 (2004).

7. J. Heikenfeld, A. Jajack, B. Feldman, S. W. Granger, S. Gaitonde, G. Begtrup, and B. A. Katchman, "Accessing analytes in biofluids for peripheral biochemical monitoring," Nature Biotechnology 37, 407-419 (2019).

8. M. Gourzi, A. Rouane, R. Guelaz, M. S. Alavi, M. B. McHugh, M. Nadi, and P. Roth, "Noninvasive glycaemia blood measurements by electromagnetic sensor: Study in static and dynamic blood circulation," Journal of Medical Engineering & Technolog 29(1), 22-26 (2005).

9. A. J. Berger, T.-W. Koo, I. Itzkan, G. Horowitz, and M. S. Feld, "Multicomponent blood analysis by near-infrared Raman spectroscopy," Applied Optics 38(13), 2916-2926 (1999).

10. J. Kottmann, U. Grob, J. Rey, and M. Sigrist, "Mid-Infrared fiber-coupled photoacoustic sensor for biomedical applications," Sensors 13(1), 535-549 (2013).

11. K. V. Larin, M. S. Eledrisi, M. Motamedi, and R. O. Esenaliev, "Noninvasive blood glucose monitoring with optical coherence tomography," Diabetes Care 25(12), 2263-2267 (2002).

12. B. H. Malik, G. L. Coté, "Real-time, closed-loop dual-wavelength optical polarimetry for glucose monitoring," Journal of Biomedical Optics 15(1), 017002 (2010).

13. X. Chen, J. Jia, and S. Dong, "Organically modified sol-gel/chitosan composite based glucose biosensor," Electroanalysis 15(7), 608-612 (2003).

14. X. Yang, L. Hua, H. Gong, and S. N. Tan, "Covalent immobilization of an enzyme (glucose oxidase) onto a carbon sol-gel silicate composite surface as a biosensing platform," Analytica Chimica Acta 478(1), 67-75 (2003).

15. J. Choi, K.-J. Kim, B. Kim, H. Lee, and S. Kim, "Covalent functionalization of epitaxial graphene by Azidotrimethylsilane," The Journal of Physical Chemistry C 113(22), 9433-9435 (2009).

16. B. Pramanick, N. Mandal, D. Mondal, C. RoyChaudhuri, and S. Chakraborty, "C-MEMS derived glassy carbon electrodes-based sensitive electrochemical biosensors," IEEE Sensors Journal 20(21), 12472-12478 (2020).

17. L. C. Clark, C. Lyons, "Electrode systems for continuous monitoring in cardiovascular surgery," Annals of the New York Academy of Sciences 102(1), 29-45 (1962).

18. W. Zhang, Y. Du, and M. L. Wang, "Noninvasive glucose monitoring using saliva nano-biosensor," Sensing and Bio-Sensing Research 4, 23-29 (2015).

19. H. Yao, Y. Liao, A. R. Lingley, A. Afanasiev, I. Lâhdesmâki, B. P. Otis, and B. A. Parviz, " A contact lens with integrated telecommunication circuit and sensors for wireless and continuous tear glucose monitoring," Journal of Micromechanics and Microengineering 22(7), 075007 (2012).

20. A. J. Bandodkar, W. Jia, C. Yardimci, X. Wang, J. Ramirez, and J. Wang, "Tattoo-based noninvasive glucose monitoring: A proof-of-concept study," Analytical Chemistry 87(1), 394-398 (2015).

21. Z. Temoçin, "Designing of a stable and selective glucose biosensor by glucose oxidase immobilization on glassy carbon electrode sensitive to H2O2 via nanofiber interface," Journal of Applied Electrochemistry 51(2), 283-293 (2021).

22. L. Zhang, S.-M. Yuan, L.-M. Yang, Z. Fang, and G.-C. Zhao, "An enzymatic glucose biosensor based on a glassy carbon electrode modified with manganese dioxide nanowires," Microchimica Acta 180, 627-633 (2013).

23. L. Zhang, Q. Chen, X. Han, and Q. Zhang, "MNO2 nanoparticles and Carbon Nanofibers Nanocomposites with high sensing performance toward glucose," Journal of Cluster Science 29(6), 1089-1098 (2018).

24. C. Kaçar, B. Dalkiran, P. E. Erden, and E. Kiliç, "An amperometric hydrogen peroxide biosensor based on Co3O4 nanoparticles and multiwalled carbon nanotube modified Glassy Carbon Electrode," Applied Surface Science 311, 139-146 (2014).

25. Q. Mei, R. Fu, Y. Ding, L. li, A. Wang, D. Duan, and D. Ye, "Electrospinning of highly dispersed ni/COO carbon nanofiber and its application in glucose electrochemical sensor," Journal of Electroanalytical Chemistry 847, 113075 (2019).

26. B. K. Shrestha, R. Ahmad, H. M. Mousa, I.-G. Kim, J. I. Kim, M. P. Neupane, C. H. Park, and C. S. Kim, "Highperformance glucose biosensor based on chitosan-glucose oxidase immobilized polypyrrole/Nafion/functionalized multiwalled carbon nanotubes bio-nanohybrid film," Journal of Colloid and Interface Science 482, 39-47 (2016).

27. H. Guan, D. Gong, Y. Song, B. Han, and N. Zhang, "Biosensor composed of integrated glucose oxidase with liposome microreactors/chitosan nanocomposite for amperometric glucose sensing," Colloids and Surfaces A: Physicochemical and Engineering Aspects 574, 260-267 (2019).

28. P. A. Gallay, M. D. Rubianes, F. A. Gutierrez, and G. A. Rivas, "Avidin and glucose oxidase-non-covalently functionalized multiwalled carbon nanotubes: A new analytical tool for building a bienzymatic glucose biosensor," Electroanalysis 31(10), 1888-1894 (2019).

29. M. Baghayeri, "Glucose sensing by a glassy carbon electrode modified with glucose oxidase and a magnetic polymeric nanocomposite," RSC Advances 5(24), 18267-18274 (2015).

30. Y. Yuan, Y. Wang, H. Wang, and S. Hou, "Gold nanoparticles decorated on single layer graphene applied for electrochemical ultrasensitive glucose biosensor," Journal of Electroanalytical Chemistry 855, 113495 (2019).

31. G. Bharath, R. Madhu, S.-M. Chen, V. Veeramani, A. Balamurugan, D. Mangalaraj, C. Viswanathan, and N. Ponpandian, "Enzymatic electrochemical glucose biosensors by Mesoporous 1D hydroxyapatite-on-2d reduced graphene oxide," Journal of Materials Chemistry B 3(7), 1360-1370 (2015).

32. Q. Zeng, J.-S. Cheng, X.-F. Liu, H.-T. Bai, and J.-H. Jiang, "Palladium nanoparticle/Chitosan-grafted graphene nanocomposites for construction of a glucose biosensor," Biosensors and Bioelectronics 26(8), 3456-3463 (2011).

33. Y. Wang, Y. Li, L. Tang, J. Lu, and J. Li, "Application of graphene-modified electrode for selective detection of dopamine," Electrochemistry Communications 11(4), 889-892 (2009).

34. X. Kang, J. Wang, H. Wu, I. A. Aksay, J. Liu, and Y. Lin, "Glucose oxidase-graphene-Chitosan modified electrode for direct electrochemistry and glucose sensing," Biosensors and Bioelectronics 25(4), 901-905 (2009).

35. C. Wang, G. Jia, L. H. Taherabadi, and M. J. Madou, "A novel method for the fabrication of high-aspect ratio C-MEMS structures," Journal of Microelectromechanical Systems 14(2), 348-358 (2005).

36. B. Pramanick, M. Vazquez-Pinon, A. Torres-Castro, S. O. Martinez-Chapaa, and M. Madou, "Effect of pyrolysis process parameters on electrical, physical, chemical and electrochemical properties of SU-8-derived carbon structures fabricated using the C-MEMS process," Materials Today: Proceedings 5(3), 9669-9682 (2018).

37. J. Huang, "Diffusion impedance of electroactive materials, electrolytic solutions and porous electrodes: Warburg impedance and beyond," Electrochimica Acta 281, 170-188 (2018).

38. I. Lelidis, G. Barbero, "Role of the displacement current on Warburg-type behavior," Physical Review E 95(5), 052604 (2017).

39. J. Huang, "Generalization of porous electrode theory for Noninteger Dimensional Space," The Journal of Physical Chemistry C 122(1), 557-565 (2018).

40. H. Li, C. Zhao, X. Wang, J. Meng, Y. Zou, S. Noreen, L. Zhao, Z. Liu, H. Ouyang, P. Tan, M. Yu, Y. Fan, Z. L. Wang, and Z. Li, "Fully bioabsorbable capacitor as an energy storage unit for Implantable Medical Electronics," Advanced Science 6(6), 1801625 (2019).

41. M. Mazaheri, A. Simchi, and H. Aashuri, "Enzymatic biosensing by covalent conjugation of enzymes to 3D-networks of graphene nanosheets on arrays of vertically aligned gold nanorods: Application to voltammetric glucose sensing," Microchimica Acta 185, 178 (2018).

42. N. Elgrishi, K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart, and J. L. Dempsey, "A practical beginner's guide to cyclic voltammetry," Journal of Chemical Education 95(2), 197-206 (2018).

i Надоели баннеры? Вы всегда можете отключить рекламу.