Научная статья на тему 'Biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis'

Biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis Текст научной статьи по специальности «Клиническая медицина»

CC BY
239
80
i Надоели баннеры? Вы всегда можете отключить рекламу.
Журнал
Хирургия позвоночника
Scopus
ВАК
Область наук
Ключевые слова
spondylolisthesis / sagittal balance / biomechanical modeling

Аннотация научной статьи по клинической медицине, автор научной работы — Aleksey Leonidovich Kudiashev, Vladimir Vasilievich Khominets, Andrey Vasilyevich Teremshonok, Evgeny Borisovich Nagorny, Sergey Yuryevich Stadnichenko

Objective. To assess the results of clinical approbation of individual finite-element biomechanical model of a patient’s spino-pelvic complex with subsequent modeling of the best option of surgical treatment. Material and Methods. A biomechanical modeling of changes in the sagittal profile of a patient with degenerative disease of the lumbosacral spine, bilateral spondylolysis, and unstable grade 2 spondylolisthesis of the L4 vertebra was performed. The developed biomechanical model made it possible to assess the characteristics of the stress-strain state of the spinal motion segments aroused due to development of the disease. Within the built biomechanical model of the patient’s spino-pelvic complex, a corrective operation was further modeled that assumed a preservation of harmonious profile of sagittal spino-pelvic relationships. Post-correction characteristics of the stress-strain state of spinal motion segments were studied and compared with preoperative parameters of the biomechanical model. Results. Using methods of biomechanics and computer modeling allowed to calculate the stress-strain state of the lumbosacral spine under static load for two options of fixation and intervertebral cage implantation at the L4–L5 level: four transpedicular screws (L4–L5 vertebrae) and six transpedicular screws (L3–L4–L5 vertebrae). The simulation results showed that neither metal implants, nor elements of the lumbosacral spine experienced critical stresses and deformations that could lead to the destruction and instability of the implant. Conclusion. The developed individual biomechanical finite-element solid model of the spine and pelvis allowed for biomechanical justification of prerequisites for the formation and further progression of degenerative changes in spinal motion segments associated with violations of the sagittal profile due to grade 2 spondylolisthesis of the L4 vertebra. The model built on the results of radiological examination biomechanically substantiated the best option of corrective spine surgery allowing to minimize stresses and deformations by choosing reasonable magnitude of correction of sagittal spino-pelvic parameters and configuration of transpedicular system.

i Надоели баннеры? Вы всегда можете отключить рекламу.

Похожие темы научных работ по клинической медицине , автор научной работы — Aleksey Leonidovich Kudiashev, Vladimir Vasilievich Khominets, Andrey Vasilyevich Teremshonok, Evgeny Borisovich Nagorny, Sergey Yuryevich Stadnichenko

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.
i Надоели баннеры? Вы всегда можете отключить рекламу.

Текст научной работы на тему «Biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis»

hirurgia pozvonochnika 2018;15(4):87-94 © a.l. kudiashev et al., 2018

biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis

A.L. Kudiashev1, V.V. Khominets1, A.V. Teremshonok1, E.B. Nagorny1, S.Yu. Stadnichenko1, A.V. Dol2, D.V. Ivanov2,

I.V. Kirillova2, L.Yu. Kossovichh2, A.L. Kovtun3

1Military Medical Academy n.a. S.M. Kirov, St. Petersburg, Russia 2Saratov National Research State University n.a. N.G. Chernyshevsky, Saratov, Russia 3Advanced Research Foundation, Moscow, Russia

Objective. To assess the results of clinical approbation of individual finite-element biomechanical model of a patient's spino-pelvic complex with subsequent modeling of the best option of surgical treatment.

Material and Methods. A biomechanical modeling of changes in the sagittal profile of a patient with degenerative disease of the lumbosacral spine, bilateral spondylolysis, and unstable grade 2 spondylolisthesis of the L4 vertebra was performed. The developed biomechanical model made it possible to assess the characteristics of the stress-strain state of the spinal motion segments aroused due to development of the disease. Within the built biomechanical model of the patient's spino-pelvic complex, a corrective operation was further modeled that assumed a preservation of harmonious profile of sagittal spino-pelvic relationships. Post-correction characteristics of the stress-strain state of spinal motion segments were studied and compared with preoperative parameters of the biomechanical model. Results. Using methods of biomechanics and computer modeling allowed to calculate the stress-strain state of the lumbosacral spine under static load for two options of fixation and intervertebral cage implantation at the L4—L5 level: four transpedicular screws (L4—L5 vertebrae) and six transpedicular screws (L3—L4—L5 vertebrae). The simulation results showed that neither metal implants, nor elements of the lumbosacral spine experienced critical stresses and deformations that could lead to the destruction and instability of the implant. Conclusion. The developed individual biomechanical finite-element solid model of the spine and pelvis allowed for biomechanical justification of prerequisites for the formation and further progression of degenerative changes in spinal motion segments associated with violations of the sagittal profile due to grade 2 spondylolisthesis of the L4 vertebra. The model built on the results of radiological examination biomechanically substantiated the best option of corrective spine surgery allowing to minimize stresses and deformations by choosing reasonable magnitude of correction of sagittal spino-pelvic parameters and configuration of transpedicular system. Key Words: spondylolisthesis, sagittal balance, biomechanical modeling.

Please cite this paper as: Kudiashev AL, Khominets VV, Teremshonok AV, Nagorny EB, Stadnichenko SYu, Dol AV, Ivanov DV, Kirillova IV, Kossovich LYu, Kovtun AL. Biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis. Hir. Pozvonoc. 2018;15(4):87—94. In Russian. DOI: http://dx.doi.org/10.14531/ss2018A.87-94.

True spondylolisthesis is the displacement of one of the lumbar vertebrae in the sagittal plane over the inferior one due to a bilateral defect in the inter-articular part of its arch. The proportion of patients with true spondylolisthesis amounts to 7-10 % of all patients complaining of persistent pain in the lumbar region [1]. According to modern concepts, the effectiveness of surgical treatment of such patients is defined not only by elimination of the compression of neural structures, but also by the required correction with the achievement of harmonious spino-pelvic relationships [2-4].

At the same time, the existing tactical approaches of orthopedic traumatolo-gists and neurosurgeons (vertebrologists)

to treatment of this category of patients are still debated to date, which results in variability in the choice of surgical technologies depending on the degree of displacement of the vertebra, segmental stability, the estimated volume of bone structure resection, and the magnitude of the required correction [4-7]. However, there are almost no studies devoted to substantiating the choice of one or another variant of surgical treatment of the specialized patients, as well as to the analysis of the achieved results from the biomechanical point of view.

This paper presents an applied bio-mechanical study aimed at substantiating the prerequisites for the development and progression of degenerative changes in the spinal motion segments

in a patient with grade 2 spondylolisthe-sis of L4 vertebra before and after surgical treatment, as well as the experience of using patient-specific biomechanical modeling of corrective spinal surgery.

The purpose of the study is to assess the results of clinical approbation of patient-specific finite-element biome-chanical model of the patient's spino-pel-vic complex with subsequent modeling of the best option of surgical treatment.

Material and Methods

At the first stage of the study, we conducted biomechanical modeling of changes in the sagittal profile of a patient with degenerative disease of the lumbosacral spine, bilateral spondylolysis,

87

unstable grade 2 spondylolisthesis of the L4 vertebra. The developed biomechanical model made it possible to assess the characteristics of the stress-strain state existing (due to the nature of the disease) in the spinal motion segments of the thoracic and lumbosacral spine. ANSYS, a finite element analysis system, was used to calculate and analyze the stresses arising in the vertebrae, intervertebral discs and implants upon application of a stationary load in different directions. Two variants of transpedicular fixation were considered: four transpedicular screws (L4-L5 vertebrae) and six transpedicular screws (L3-L4-L5 vertebrae).

At the second stage, the constructed patient-specific biomechanical model of the spine was used to simulate corrective surgery. The sagittal spino-pelvic parameters were corrected until a harmonious profile was achieved. We have subsequently re-examined the characteristics of the stress-strain state of the spinal motion segments of the thoracic and lumbar spine and calculated the magnitude and localization of stresses arising in the vertebrae, intervertebral discs and facet joint in case of standard load. We compared the obtained results to similar parameters at the first stage of the study.

Initial data. The results of CT scans of all parts of the spine and pelvis (from the level of C7 vertebra to the proximal parts of the femurs), as well as full body X-ray performed with the patient standing in two projections.

Creating a 3D-model of the spino-pel-vic complex. At the initial stage, a three-dimensional computer model of the spine was built on the basis of CT data. Then, three-dimensional models of fixation systems (cage at the level of L4-L5) and transpedicular screws (4 screws for fixing L4-L5 vertebrae and 6 screws for fixing L3-L4-L5 vertebrae) were created. Then the models of the fixation systems and the spine were combined to produce the models shown in Fig. 1.

Finite-element modeling. The ANSYS 18 finite-element analysis system was used for numerical simulation, in which the stresses in the vertebrae, intervertebral discs and fixation systems were

calculated and analyzed. The spinal loading was simulated by bending and twisting moments in three anatomical planes, which were applied to the upper endplate of the L1 vertebra (Fig. 2, red arrows) [8, 9]. The magnitude of the moments was 7.5 N • m [10].

Thus, we calculated two types of operations with four loading options. Mechanical characteristics of the elements of the spine and the implants were taken from literary sources [8, 9, 11]. The sacrum was firmly secured in three directions.

Results and Discussion

The results of finite-element modeling are presented for two variants of spinal fixation at the levels L4-L5 and L3-L4-L5 with the installation of a PEEK-ceramic cage at the L4-L5 level.

Fig. 3 shows the biomechanical fields of movement for the lumbosacral spinal segments in the case of loading with a twisting moment. The displacement field is typical for other investigated loading options.

The largest values of displacements for both types of fixation and four applied loading moments are shown in Table 1.

With all the load cases considered, the four-screw design turned out to be more rigid and stable than the six-screw design. This conclusion is confirmed by the third column in Table 1; the fourth column shows the differences in fixation with four and six screws. Since the 4-screw fixation does not immobilize L3 vertebra, its range of movements is wider in case of 4-screw fixation.

The analysis of the deformations in the discs (Table 2) shows that the fixation option with four screws is preferable from the biomechanical point of view. In fact, the deformations of the L5-S1, L2-L3 and L1-L2 discs with 4-screw fixation are not larger than those of the same discs when the metal structure is installed using six pedicle screws. And only the L3-L4 disc turns out to be more deformed in the case of 4-screw fixation, since the 6-screw system provides better fixation to it.

Table 3 shows the highest values of effective stresses arising in the lumbar vertebrae for both models under review.

In the case of the 6-screw configuration of the transpedicular fixators, the stress in L1 and L2 vertebrae turned out to be the same as for the 4-screw configuration. The 6-screw configuration resulted in higher stress in the L3 vertebra than the 4-screw configuration of the fixation system. For L4 and L5 vertebrae, the stress in case of the 4-screw configuration is higher than with the 6-screw configuration, but the orders of magnitude were the same.

Therefore, it can be concluded that from the biomechanical point of view the transpedicular fixation at a single level (with the 4-screw system) is preferable to fixation at two levels, since it provides higher stability and does not load intervertebral discs above and below the fixed segment.

Patient N., 19 years old, who suffered from the degenerative disease of the lum-bosacral spine, bilateral spondylolysis, unstable grade 2 L4 spondylolisthesis and pain vertebrogenic and muscular tonic syndrome underwent L4 laminectomy, L4-L5 discectomy, posterior internal correction (reduction) and fixation of the spine with transpedicular system at the level of L4-L5, posterior interbody spinal fusion with a cage, the L4-L5 posterolateral fusion with autografts (Fig. 4) in accordance with the preoperative planning and the performed biomechanical modeling.

Conducting the surgery based on the data of the performed biomechanical modeling and planning made it possible to achieve the calculated harmonious values of the sagittal spino-pelvic relationships (Table 4).

Conclusion

From the biomechanical point of view, fixation with four transpedicular screws is preferable to fixation with six screws, since it is more stable and is characterized by lower stresses and deformations in the vertebrae and intervertebral discs. Moreover, this variant of the surgery does not require the fixation

88

Fig. 1

X-ray of the spine (sagittal projection) of a patient with unstable grade 2 spondylolisthesis of L4 vertebra (a) and models of the lumbosacral spine with established pedicle fixation systems: 4 screws on the left, 6 screws on the right (b)

of the L3-L4 spinal motion segment that is not subject to degenerative changes, which would have entailed a cascade of degenerative changes in the corresponding intervertebral disc. The correction achieved with the use of this arrangement of the transpedicular system in combination with anterior

fusion is sufficient not only for the reduction of L4 vertebra, but also for balanced spino-pelvic relationships.

Achievement of the harmonious sagittal profile (correction of the deformity) in combination with optimal fixation of the spine from the biological and bio-mechanical points of view ensured that

the spino-pelvic complex is in a state in which stresses and deformations are minimized in the structures of the spinal motion segments, in the pedicle screws, rods and interbody cage.

Therefore, patient-specific biomechanical modeling of available options for correction and fixation of the spine has demonstrated its effectiveness in clinical practice for simulating the bio-mechanical parameters of the functioning of spinal segments in the postoperative period.

The study did not have sponsorship. The authors declare no conflict of interest.

Fig. 2

Applied twisting and bending moments

89

Loading moment Type of fixation L4—L5 L3—L4—L5

Table 1

Movement in the elements of the lumbosacral spine, mm

4 screws

6 screws

4 screws

6 screws

4 screws

6 screws

4 screws

6 screws

1.00

1.30

0.75

0.87

1.00

1.30

0.70

0.80

2.00

2.00

2.30

1.20

2.00

1.80

1.90

1.10

90

_hirurgia pozvonochnika 2018;15(4):87-94_

a.l. kudiashev et al. biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis

Table 2

Deformations in the intervertebral discs, mm

Loading moment Type L5—S1 L3-L4 L2-L3 L1—L2

of fixation

4 screws 0.08 0.09 0.08 0.10

6 screws 0.12 0.05 0.09 0.10

4 screws 0.23 0.20 0.16 0.22

6 screws 0.30 0.01 0.19 0.39

4 screws 0.08 0.09 0.08 0.10

6 screws 0.12 0.05 0.09 0.10

4 screws 0.18 0.13 0.08 0.09

6 screws 0.25 0.01 0.08 0.11

91

Table 3

Tension in the vertebrae, MPa

Loading moment Type of fixation L1 L2 L3 L4 L5

4 screws

6 screws

4 screws

6 screws

4 screws

6 screws

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

4 screws

6 screws

4 4

Table 4

Characteristics of the patient's sagittal spino-pelvic parameters, deg.

Parameters

Prior to the surgery After the surgery

Pelvic Incidence Sacral Slope Pelvic Tilt L1—S1 L4—S1

47 27

20 51 18

47

36 15 61 35

4 46

49

4 4 57 43 58

6 7 5 38 76

6 7 45 24 35

4 4 4 46 49

4 4 57 43 59

4 4 5 45 44

4 4 50 33 27

Calculated

47 39 ± 6 9 ± 6 63 ± 11 42 ± 7

Fig. 4

X-ray of the spine of patient N., 19 years old, with unstable grade 2 spondylolisthesis of L4 vertebra after corrective surgery (sagittal projection)

92

References

1. Vialle R, Ilharreborde B, Dauzac C, Lenoir T, Rillardon L, Guigui P. Is there a sagittal imbalance of the spine in isthmic spondylolisthesis? A correlation study. Eur Spine J. 2007;16:1641-1649. DOI: 10.1007/s00586-007-0348-4.

2. Dubousset J. Basic principles of spine surgery. Hir. Pozvonoc. 2016;13(4):95-103. In Russian. DOI: https://doi.org/10.14531/ss20164.95-103.

3. Krutko AV. Sagittal Balance. Harmony in Formulas. Novosibirsk, 2016. In Russian.

4. Le Huec JC, Faundez A, Domínguez D, Hoffmeyer P, Aunoble S. Evidence showing the relationship between sagittal balance and clinical outcomes in surgical treatment of degenerative spinal diseases: a literature review. Int Orthop. 2015;39:87-95. DOI: 10.1007/s00264-014-2516-6.

5. Dubousset J. The spine is three-dimensional entity, though 3d alignment and 3d balance should not be confused. Hir. Pozvonoc. 2016;13(2):77-85. In Russian. DOI: https://doi.org/10.14531/ss2016.277-85.

6. /link DM, Nikulina AA. Developing a system for efficient analysis of lumbosacral sagittal balance. Science and Education of the Bauman MSTU. Electronic journal. 2015;(6):346-360. In Russian. DOI: http://dx.doi.org/10.7463/0615.0783321.

7. Kaneko K, Aota Y, Sekiya T, Yamada K, Saito T. Validation study of arm positions for evaluation of global spinal balance in EOS imaging. Eur J Orthop Surg. Traumatol. 2016;7:725-733. DOI: 10.1007/s00590-016-1813-8.

8. Dreischarf M, /ander T, Shirazi-Adl A, Puttlitz CM, Adam CJ, Chen CS, Goel VK, Kiapour A, Kim YH, Labus KM, Little JP, Park WM, Wang YH, Wilke HJ, Rohlmann A, Schmidt H. Comparison of eight published static finite element mod-

els of the intact lumbar spine: predictive power of models improves when combined together. J Biomech. 2014;47:1757-1766. DOI: 10.1016/j.jbiomech.2014.04.002.

9. Wang J, Zhong ZC, Cheng CK, Chen CS, Yu CH, Chang TK, Wei SH.

Finite element analysis of the spondylolysis in lumbar spine. Biomed Mater Eng. 2006;16:301-308.

10. Wilke HJ, Wenger K, Claes L. Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J. 1998;7:148-154. DOI: 10.1007/s005860050045.

11. Ben-Hatira F, Saidane K, Mrabet A. A finite element modeling of the human lumbar unit including the spinal cord. J Biomedical Science and Engineering. 2012;5:146-152. DOI: 10.4236/jbise.2012.53019.

Address correspondence to:

Kudiashev Aleksey Leonidovich Military Medical Academy n.a. S.M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, a.kudyashev@gmail.com

Received 13.07.2018 Review completed07.09.2018 Passed for printing 11.09.2018

Aleksey Leonidovich Kudiashev, MD, PhD, associate professor, deputy head of the Department and Clinic of Military Traumatology and Orthopedics of Military Medical Academy n.a. S.M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, a.kudyashev@gmail.com;

Vladimir Vasilievich Khominets, DMSc, Professor, Head of the Department and Clinic of Military Traumatology and Orthopedics of Military Medical Academy n.a. S.M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, khominets_62@mail.ru;

Andrey Vasilyevich Teremshonok, MD, PhD, assistant professor of the Department of Military Traumatology and Orthopedics, Military Medical Academy n. a. S. M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, Teremshonok@rambler.ru;

Evgeny Borisovich Nagorny, MD, PhD, Associate Professor, Lecturer of the Department of Military Traumatology and Orthopedics of Military Medical Academy n.a. S.M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, polartravma@rambler.ru;

Sergey Yuryevich Stadnichenko, clinical resident of the Department of Military Traumatology and Orthopedics of Military Medical Academy n.a. S.M. Kirov, Academika Lebedeva str., 6, St. Petersburg, 194044, Russia, s.stadnichenko@inbox.ru;

Aleksandr Viktorovich Dol, PhD in Physics and Mathematics, senior researcher of the Laboratory of Clinical Decision Support Systems, Saratov State University, Astra-khanskaya str., 83, Saratov, 410012, Russia, cdss@sgu.ru;

Dmitry Valeryevich Ivanov, PhD in Physics and Mathematics, leading researcher of the Laboratory of Clinical Decision Support Systems, Saratov State University, Astrakhanskaya str., 83, Saratov, 410012, Russia, cdss@sgu.ru;

Irina Vasilyevna Kirillova, PhD in Physics and Mathematics, associate professor, Head of the Laboratory of Clinical Decision Support Systems, Saratov State University, Astrakhanskaya str., 83, Saratov, 410012, Russia, cdss@sgu.ru;

Leonid Yuryevich Kossovich, Doctor of Physics and Mathematics, Processor, Scientific Supervisor of the Laboratory of Clinical Decision Support Systems, Saratov State University, Astrakhanskaya str., 83, Saratov, 410012, Russia, prezident@sgu.ru;

Anatoly Leonidovich Kovtun, Doctor of Biological Sciences, Professor, Head of the Project Group of chemical-biological and medical research direction of the Advanced Research Foundation, Berezhkovskaya nab., 22, building 3, Moscow, 121059, Russia, KovtunAL@fpi.gov.ru.

93

_hirurgia pozvonochnika 2018;15(4):87-94_

a.l. kudiashev et al. biomechanical modeling in surgical treatment of a patient with true lumbar spondylolisthesis

_94_

degenerative diseases of the spine

i Надоели баннеры? Вы всегда можете отключить рекламу.