Научная статья на тему 'THE EFFECT OF NOISE IN RAMAN SPECTRA ON THE RECONSTRUCTION OF THE CONCENTRATION OF AMINO ACIDS IN THE MIXTURE BY MULTIVARIATE CURVE RESOLUTION (MCR) ANALYSIS'

THE EFFECT OF NOISE IN RAMAN SPECTRA ON THE RECONSTRUCTION OF THE CONCENTRATION OF AMINO ACIDS IN THE MIXTURE BY MULTIVARIATE CURVE RESOLUTION (MCR) ANALYSIS Текст научной статьи по специальности «Медицинские технологии»

CC BY
22
7
i Надоели баннеры? Вы всегда можете отключить рекламу.
Ключевые слова
SIGNAL-TO-NOISE RATIO / RAMAN SPECTRUM / RAMAN SCATTERING / MULTIVARIATE CURVE RESOLUTION / FREE AMINO ACIDS

Аннотация научной статьи по медицинским технологиям, автор научной работы — Matveeva Irina A., Bratchenko Lyudmila A., Myakinin Oleg O., Tupikova Elena N., Zakharov Valery P.

Changes in the concentration of free amino acids in biological tissues is a sign of impaired protein metabolism in patients with cancer. Recently, Raman spectroscopy has been used for early diagnostics of oncological diseases. The concentrations of individual components of biological tissue (for instance, the concentrations of amino acids) can be obtained by decomposing the tissue Raman spectrum. This study was designed to evaluate the effect of noise in the Raman spectra of individual amino acids on the result of the decomposition of the spectra of an amino acid mixture. As a decomposition method, we used Multivariate Curve Resolution-Alternating Least Squares (MCR-ALS) analysis and investigate experimental Raman spectra of amino acids and mathematically simulated Raman spectra of amino acid mixtures. Noise with different signal-to-noise ratios (SNR) was artificially added to both the experimental spectra of pure amino acids and the spectra of the mixtures. Concentration values for each amino acid obtained as a result of applying the MCR-ALS analysis have been compared with the corresponding true values and the correlation coefficients have been calculated. The results show a less pronounced negative effect of noise in the case when the spectra of pure amino acids (which were used as a basis for the MCR-ALS analysis) are noisy, and a more pronounced negative effect when the spectrum of the mixture is noisy. The accuracy of reconstruction of an amino acid is also negatively affected by strong background fluorescence in the amino acid spectrum. Moreover, the results indicate that using the basis spectra with a high SNR (SNR = 5) makes it possible to successfully estimate the amino acid concentrations in a mixture even when the Raman spectrum of the mixture is noisy and has a low SNR (SNR < 5).

i Надоели баннеры? Вы всегда можете отключить рекламу.
iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.
i Надоели баннеры? Вы всегда можете отключить рекламу.

Текст научной работы на тему «THE EFFECT OF NOISE IN RAMAN SPECTRA ON THE RECONSTRUCTION OF THE CONCENTRATION OF AMINO ACIDS IN THE MIXTURE BY MULTIVARIATE CURVE RESOLUTION (MCR) ANALYSIS»

The Effect of Noise in Raman Spectra on the Reconstruction of the Concentration of Amino Acids in the Mixture by Multivariate Curve Resolution (MCR) Analysis

Irina A. Matveeva*, Lyudmila A. Bratchenko, Oleg O. Myakinin, Elena N. Tupikova, and Valery P. Zakharov

Samara National Research University, 34 Moskovskoe shosse, Samara 443086, Russia * e-mail: m-irene-a@yandex.ru

Abstract. Changes in the concentration of free amino acids in biological tissues is a sign of impaired protein metabolism in patients with cancer. Recently, Raman spectroscopy has been used for early diagnostics of oncological diseases. The concentrations of individual components of biological tissue (for instance, the concentrations of amino acids) can be obtained by decomposing the tissue Raman spectrum. This study was designed to evaluate the effect of noise in the Raman spectra of individual amino acids on the result of the decomposition of the spectra of an amino acid mixture. As a decomposition method, we used Multivariate Curve Resolution-Alternating Least Squares (MCR-ALS) analysis and investigate experimental Raman spectra of amino acids and mathematically simulated Raman spectra of amino acid mixtures. Noise with different signal-to-noise ratios (SNR) was artificially added to both the experimental spectra of pure amino acids and the spectra of the mixtures. Concentration values for each amino acid obtained as a result of applying the MCR-ALS analysis have been compared with the corresponding true values and the correlation coefficients have been calculated. The results show a less pronounced negative effect of noise in the case when the spectra of pure amino acids (which were used as a basis for the MCR-ALS analysis) are noisy, and a more pronounced negative effect when the spectrum of the mixture is noisy. The accuracy of reconstruction of an amino acid is also negatively affected by strong background fluorescence in the amino acid spectrum. Moreover, the results indicate that using the basis spectra with a high SNR (SNR = 5) makes it possible to successfully estimate the amino acid concentrations in a mixture even when the Raman spectrum of the mixture is noisy and has a low SNR (SNR < 5). © 2021 Journal of Biomedical Photonics & Engineering.

Keywords: Signal-to-noise ratio; Raman spectrum; Raman scattering; multivariate curve resolution; free amino acids.

Paper #3424 received 23 May 2021; revised manuscript received 25 Jun 2021; accepted for publication 26 Jun 2021; published online 30 Jun 2021. doi: 10.18287/JBPE21.07.020309.

1 Introduction

Cancer is the second leading cause of death worldwide. The most common type of cancer was lung cancer. According to the World Health Organization, there was about 2.21 million new cases in 2020. Moreover, lung

cancer was the most common cause of cancer death in 2020 (1.80 million deaths). It is well known that lung cancer risk and mortality can be reduced by early detection of cases [1].

It is also known that during the malignant tumor development, changes in protein metabolism occur in the

patient's body. Many researchers have described changes in plasma free amino acid (PFAA) profiles in patients with cancer. Kubota, Meguid, and Hitch [2] analyzing PFAA in venous blood of patients with breast cancer, gastrointestinal tract cancer, and head and neck cancer suggested that PFAA profiles correlate diagnostically with the organ-site origin of three different kinds of malignant tumors. Miyagi et al. [3] determined the characteristics of the PFAA profiles in cancer patients with one of five types of cancer: lung, gastric, colorectal, breast, or prostate cancer. PFAA profiling for detecting lung cancer was also studied by Shingyoji et al. [4], Zhao et al. [5], and Proenza et al. [6]. These findings suggest that PFAA profiling has great potential for improving early detection of lung cancer.

It makes us look for new methods to analyze PFAA profiles. Recently, Raman spectroscopy has been used for early diagnostics of oncological diseases. Bratchenko et al. [7] have shown that this method can be used in the diagnosis of cancer, such as skin neoplasms. Moreover, Raman spectroscopy is an optical method that is relevant for the analysis of liquid media. So, we believe that Raman spectroscopy can be used for non-invasive analysis of blood plasma. The Raman spectra of PFAAs are specific and can be used to successfully evaluate PFAAs' concentration in a mixture of different substances by the Raman spectrum of this mixture.

In this study, we use Multivariate Curve Resolution-Alternating Least Squares (MCR-ALS) method to analyze the Raman spectra of amino acid mixtures. This method is widely used to reconstruct the concentration profiles of chemicals analysis [8, 9]. Recently, MCR-ALS method has found wide biological and medical applications [8, 10] and has been used for the analysis of spectral data when it is required to determine the concentrations of complex mixture components from spectra. For example, Xu and Rice [11] used a MCR spectral unmixing in fluorescence imaging. Chen et al. [12] used Raman spectroscopic detection of keratin with MCR analysis for automatic oral cancer diagnosis. Iwasaki et al. [13] investigated the possibilities of discrimination of breast cancer cells from normal mammary epithelial cells by Raman microspectroscopy and MCR analysis. It should be noted that the use of MCR-ALS analysis makes it possible not only to estimate the concentrations of components, but also to obtain their "pure" Raman spectra [14].

However, the MCR-ALS analysis can be sensitive to noise in the Raman spectra analyzed. In the practical application, the efficiency of evaluation of PFAAs' concentration in a mixture may decrease due to the fact that the Raman spectra contain a noise signal. We also suppose that, if MCR-ALS uses a known predetermined basis of amino acids, the result can be affected by noise in both the pure amino acid Raman spectra of the basis and Raman spectra of the mixtures. On the one hand, we can provide high-quality registration of basis Raman spectra (that is, the spectra of pure amino acids that are used in the MCR-ALS analysis) with high signal-to-noise ratio (SNR) using a spectroscopic setup with high

spectral resolution and increasing integration time. In addition, pure amino acids may be available for registration (or Raman microscopy of the samples may be used). And finally, it is enough to record the spectra (which are used as the basis) only once - then one can use them when analyzing other samples of mixtures. On the other hand, in a clinical setting, high-quality recording of Raman spectra can be difficult due to the large patient flow and limited time to examine a patient. Therefore, it can be assumed that the Raman spectra, which are subject to analysis by the MCR-ALS method, will have a lower SNR than the previously obtained Raman spectra of amino acids used as a basis for the MCR-ALS analysis.

This study was designed to evaluate an effect of the noise in Raman basis spectra of amino acids and Raman spectra of a mixture on their reconstruction from the mixture.

2 Materials and Methods

The experimental Raman spectra used in our study are recorded using a portable spectroscopic setup which includes a thermally stabilized LML-785.0RB-04 laser diode module as an excitation source (785 ± 0.1 nm central wavelength, 200 mW laser power) and a QE 65 Pro spectrometer (OceanOptics, Inc., USA) with CCD detector operating at -15 °C [15].

We used 20 standard proteinogenic amino acids (see Table 1 and Fig. 1). Amino acids are presented in crystalline powder form. The amino acids in the crystal form were placed onto the metal-coated slide. All spectra were registered at the room temperature. The registration of spectra using this system was carried out in 800-1000 nm with 0.2 nm spectral resolution that corresponds to the 240-2236 cm-1. The Raman signal of amino acids was acquired from 3 accumulations each of 5 sec integration time. The mean Raman spectrum for each amino acid was averaged over three registered spectra.

The preprocessing of the registered data includes only cosmic ray and dark noise removal that are automatically applied in the "Spectra Suite" software package (OceanOptics, Inc., USA) [16]. Examples of recorded and preprocessed spectra is shown in Fig. 1.

Then we have modeled amino acid mixtures. The "mixture" means the sum of 20 standard proteinogenic amino acids (see Table 1), taken in different quantities, that is, with different concentrations. The concentration of amino acids in the mixtures is chosen so that the mixtures correspond to real PFAA profiles of blood plasma samples studied by other researchers [3-6].

We define a Raman spectrum of an amino acid mixture as a mathematical sum of the products of the pure amino acid spectra of and its concentrations in the mixture:

SM (l) = CA1 X SA1

(X) + - + CAN X S AN (l), (1)

Table 1 The standard proteinogenic amino acids. Amino acid Glycine Alanine Valine Isoleucine

Letter code Gly Ala Val Ile

Formula C2H3NO C3H5NO C5H9NO C6H11NO

Skeleton formula "-"Of jy

Amino acid Leucine Serine Threonine Aspartic acid

Letter code Leu Ser Thr Asp

Formula C6H11NO C3H5NO2 C4H7NO2 C4H5NO3

Skeleton formula

Amino acid Glutamic acid

Asparagine

Glutamine

Lysine

Letter code Glu Asn Gln Lys

Formula C5H7NO3 C4H6N2O2 C5H8N2O2 C6H12N2O

Skeleton formula ¿f

Amino acid Arginine Cysteine Methionine Phenylalanine

Letter code Arg Cys Met Phe

Formula C6H12N4O C3H5NOS C5H9NOS C9H9NO

Skeleton formula

Amino acid Tyrosine

Skeleton formula

Tryptophan

Histidine

Proline

Letter code Tyr Trp His Pro

Formula C9H9NO2 C11H10N2O C6H7N3O C5H7NO

Glycine (Gly)

Alanine (Ala)

Valine (Val)

Isoleucine (lie)

fiBiaJJ t^kklj

1250 1750

Raman shift, cm-1 Leucine (Leu)

750 1250 1750

Raman shift, cm1 Serine (Ser)

750 1250 1750

Raman shift, cm1 Threonine (Thr)

250 750 1250 1750 2250

Raman shift, cm1 Aspartic acid (Asp)

i ^ "T

í:ikW-_ fttra

250 750 1250 1750 2250

Raman shift, cm-1 Glutamic acid (Glu)

750 1250 1750

Raman shift, cm"1 Asparagine (Asn)

250 750 1250 17Í

Raman shift, cnr1 Glutamine (Gin)

250 750 1250 1750 2250

Raman shift, cm" Lysine (Lys)

fro^,

250 750 1250 1750 2250

Raman shift, cm-1 Arginine (Arg)

r t4jJaaJ

750 1250 1750

Raman shift, cm-1 Cystein (Cys)

750 1250 1750

Raman shift, cnr1 Methionine (Met)

250 750 1250 1750 2250

Raman shift, cm-1

Phenylalanine (Phe)

I

250 750 1250 1750 2250

Raman shift, cm4

Tyrosine (Tyr)

750 1250 1750

Raman shift, cm-1 Tryptophan (Trp)

750 1250 1750

Raman shift, cm-1 Histidine (His)

250 750 1250 1750 2250

Raman shift, cnr1

Proline (Pro)

fruí

250 750 1250 1750 2250

Raman shift, cm-1

750 1250 1750 Raman shift, cm-1

750 1250 1750 2250 Raman shift, cm-1

250 750 1250 1750 2250

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

Raman shift, cm1

Fig. 1 Recorded and preprocessed Raman spectra of the amino acids.

where cA1,...,cM - are the concentrations of pure amino acids in the mixture, sAl (l),...,sAN (l) - are the pure

amino acid spectra.

Using information on PFAA profiles in lung cancer patients [3-6], we have artificially modelled 10 Raman spectra: 5 spectra of lung cancer patient PFAA profiles and 5 spectra of control group PFAA profiles. Concentrations of amino acids in the mixtures are presented in Table 2.

The next step in our study was the simulation of a noise in the Raman spectra. It should be noted that the spectra recorded have a noise signal due to the spectroscopic setup. However, in this study, we investigate the effect of additively added noise. Therefore, the noise contained in the Raman spectra initially is not taken into account.

We added the noise as the random value process with hypothetically normal distribution, zero mean value and various standard deviations. In this case, the formula for a Raman spectrum looks like this:

where S(l) is an original Raman spectrum, sE (l) is

the a noise spectrum.

To simulate different noise levels, we evaluated an additive noise level from a signal-to-noise ratio (SNR) metric proposed in [15]:

SNR

S

6 xs

(3)

where S is the Raman signal level (intensity of Raman peak in 980-1025 cm-1 band);^ is the noise standard

SE

deviation.

To investigate the effect of the noise, it is necessary to compare different combinations of a noisy mixture and noisy basis spectra. For this purpose, we simulated Raman spectra of pure amino acids with SNRs equal to 1, 5, and 10 and Raman spectra of mixtures with SNRs equal to 2, 3, 4, and 5. Examples of the noisy spectra are shown in Fig. 2.

S' (X) = S (X) + (X)

(2)

E

Table 2 Concentrations of amino acids in the mixtures.

Concentrations of amino acids, ^M

Amino

Lung cancer patients

Control patients

№1 [4] №2 [4] №3 [5] №4 [3] №5 [6] №1 [4] №2 [4] №3 [5] №4 [3] №5 [6

Gly 214.3 208 298 220.7 133 209.7 203.5 454 207.2 120

Ala 324.2 366.5 643 356.1 213 343.1 353.5 611 348.1 231

Val 215.1 242.1 241 225.8 140 220.5 230 281 222.4 152

Ile 64.7 73.9 158 66.6 35.3 60.8 64.4 190 61.6 36.1

Leu 117.6 135.2 188 123.8 83.5 118.8 125.2 166 120.1 87.6

Ser 107.8 110.7 253 113.6 68.1 108.8 106.7 258 108.5 60.8

Thr 115.8 122.6 270 122.8 68.8 118.5 121.8 214 119.8 74.7

Asp 0 0 105 0 8.6 0 0 54 0 5.8

Glu 0 0 211 0 39.7 0 0 210 0 37.1

Asn 42.6 47.9 0 44.7 31.6 45.2 46.1 0 45.7 24

Gln 547.4 577.5 0 555.3 355 586.9 587.1 0 585.3 360

Lys 183.9 195.7 325 192.9 78.2 189.1 191.6 279 189.8 83.1

Arg 93.1 100.4 101 96.4 0 95.1 96.4 96 95.4 0

Cys 0 0 84 0 0 0 0 68 0 0

Met 24.2 26.9 57 25.6 16.1 25.8 26.8 61 26.1 14.8

Phe 59.5 66.5 139 61.3 40.3 59.6 61.3 109 59.5 35.4

Tyr 65.9 71.9 74 67.4 45.3 65.2 67.3 71 65.5 43.2

Trp 51.3 56.4 53 54.2 0 57 59.6 298 57.3 0

His 69.7 74.1 238 74.1 34.7 69.7 81.2 184 80.3 38.4

Pro 141.6 157.3 129 151.1 99.3 132.3 138.6 339 135.1 121

Fig. 2 Raman spectrum of alanine with different SNRs.

For unmixing spectra by MCR-ALS analysis we used a protocol by Felten et al. [14]. The main idea of MCR-ALS is to decompose the Raman spectra matrix D into smaller matrices C and ST:

D = C x ST + E, (4)

where C represents the concentration profiles for each of the amino acids, SP is the pure amino acid spectra matrix, and E is the error matrix.

As a basis (S), we used the noisy amino acid spectra of 20 standard proteinogenic amino acids (see Table 1) to which we added noise in the previous step. After initial estimation is given for C, it is optimized iteratively using an alternative least squares algorithm (ALS) until convergence is reached [14].

3 Results and Discussion

In this study, we investigated the following combinations of a mixture and basis spectra:

- no noise in mixture spectra, no noise in basis spectra;

- no noise in mixture spectra, SNR = 10 for basis spectra;

- no noise in mixture spectra, SNR = 5 for basis spectra;

- no noise in mixture spectra, SNR = 1 for basis spectra;

- SNR = 10 for mixture spectra, no noise in basis spectra [17];

Table 3 Correlation coefficients between obtained and true amino acid concentration values for the case of the noise added to basis spectra only.

SNR

Amino acids -

_¥ (no noise)_10_5_1_

Gly 1.000 1.000 1.000 0.996

Ala 1.000 1.000 1.000 0.999

Val 1.000 1.000 1.000 0.998

Ile 1.000 0.999 1.000 0.988

Leu 1.000 0.997 0.999 0.908

Ser 1.000 0.999 1.000 0.994

Thr 1.000 1.000 1.000 0.999

Asp 1.000 0.997 1.000 0.996

Glu 1.000 1.000 1.000 1.000

Asn 1.000 1.000 1.000 0.999

Gln 1.000 1.000 1.000 1.000

Lys 1.000 1.000 0.999 0.982

Arg 1.000 1.000 1.000 0.994

Cys 1.000 0.987 0.999 0.998

Met 1.000 1.000 1.000 0.998

Phe 1.000 1.000 1.000 1.000

Tyr 1.000 0.988 0.979 0.989

Trp 1.000 0.999 0.999 0.996

His 1.000 1.000 1.000 0.998

Pro 1.000 0.999 1.000 1.000

- SNR = 5 for mixture spectra, no noise in basis

spectra [17];

- SNR = 1 for mixture spectra, no noise in basis

spectra [17];

- SNR = 5 for mixture spectra, SNR = 5 for basis

spectra;

- SNR = 4 for mixture spectra, SNR = 5 for basis

spectra;

- SNR = 3 for mixture spectra, SNR = 5 for basis

spectra;

- SNR = 2 for mixture spectra, SNR = 5 for basis

spectra.!

As a result of the MCR-ALS analysis, we have

obtained a matrix of amino acid concentrations in the mixture spectra. Concentrations for each amino acid have been compared with the corresponding true values and the correlation coefficients have been calculated between the true concentration array and the obtained concentration array (see Tables 3, 4, 5).

Each concentration array corresponds to one of the amino acids and different mixtures (all the mixtures which we investigated). That is, each element of the array is a concentration of an amino acid in a mixture spectrum (a spectrum of one of the mixtures we used). The correlation coefficient indicates the degree of linear relation of the arrays and varies in the range between -1 and +1, where zero value corresponds to completely uncorrelated arrays.

Table 4 Correlation coefficients between obtained and true amino acid concentration values for the case of the noise added to mixture spectra only [17].

Amino acids ¥ (no noise) 10 5 1

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

Gly 1.000 0.999 0.991 0.966

Ala 1.000 0.999 0.995 0.959

Val 1.000 0.997 0.989 0.934

Ile 1.000 0.991 0.977 0.760

Leu 1.000 0.983 0.883 0.139

Ser 1.000 0.987 0.773 0.547

Thr 1.000 0.998 0.997 0.988

Asp 1.000 0.992 0.988 0.661

Glu 1.000 1.000 0.999 0.943

Asn 1.000 0.999 0.998 0.984

Gln 1.000 1.000 0.999 0.979

Lys 1.000 0.998 0.964 0.703

Arg 1.000 0.998 0.982 0.846

Cys 1.000 0.968 0.948 0.599

Met 1.000 0.997 0.953 -0.005

Phe 1.000 1.000 1.000 0.994

Tyr 1.000 0.923 0.448 0.344

Trp 1.000 0.991 0.966 0.641

His 1.000 0.999 0.990 0.901

Pro 1.000 0.998 0.992 0.899

Table 3 shows the correlation coefficients for the case when the basis spectra are noisy with a different SNR, and the mixtures have no noise. For comparison, Table 4 shows the results that we obtained in the previous study, when we investigated the effect of noise in the mixture Raman spectra on the quality of unmixing spectra [17].

As one can see from Table 3, the correlation coefficient between true and reconstructed concentrations of amino acids equals 1 for the case without noise. In the case of noisy Raman spectra, the quality of reconstruction of amino acid concentrations is expected to decrease. Nevertheless, in the cases of SNR from 10 to 1, the correlation coefficients are high for all amino acids and ranges from 0.90 to 1.

Paying attention to the comparison of these results with those obtained in the previous study (see Table 4), one can see the following. While in the case of a noise in the basis spectra, amino acids were reconstructed successfully with any of the considered SNRs, in the case of a noise in the mixture spectrum we failed to reconstruct all amino acids when SNR is less than 10. So, when SNR = 5, three out of 20 amino acids are reconstructed with an accuracy of less than 90%, and in the case of SNR = 1 there are already 11 such amino acids.

The results of our previous study [17] show that the concentrations of amino acids are restored with lower correlation coefficients if their Raman spectra have no intense peaks. That is, the ratio between Raman peaks and background, apparently caused by fluorescence, is

not high (compare the Raman spectra of cysteine, tyrosine, tryptophan with weak background fluorescence and the Raman spectra of glutamic acid, methionine, phenylalanine with strong background fluorescence in Fig. 1).

It can be concluded that a noise in the Raman spectrum of an amino acid mixture decreases the quality of reconstruction more than a noise in the basis, which is used in MCR-ALS. This may be due to the random nature of the noise we added to the spectra. It has zero mean value; therefore, during MCR-ALS analysis (see Eq. 4), the noise components compensate each other.

Table 5 demonstrates the correlation coefficients of reconstruction of mixture Raman spectra with different SNR using noisy basis Raman spectra of amino acids (SNR = 5). As expected, at higher mixture SNR (SNR > 4) the correlation coefficients are in a range from 0.9 to 1 for almost all amino acids except cysteine and tyrosine, which are characterized by low Raman peaks in all spectral range (see Fig. 1). For noisier mixture Raman spectra with SNR = 3 and SNR = 2 MCR-ALS method failed to reconstruct 5 and 6 amino acids with high quality, respectively. It should be noted that one of the failed components is cysteine, which either is absent in mixtures analyzed or its concentration is low.

As in our previous study [17], the concentrations of leucine, serine, cysteine, tyrosine, tryptophan are restored with lower correlation coefficients because random noise overlaps their spectra (which have a low ratio between Raman peaks and background fluorescence).

Table 5 Correlation coefficients between obtained and true amino acid concentration values for the case of the noise added to both basis spectra (SNR = 5 for all the cases) and mixture spectra.

SNR of mixture Raman spectra

Amino acids 5 4 3 2

Gly 0.996 0.995 0.988 0.991

Ala 0.994 0.990 0.988 0.956

Val 0.986 0.991 0.985 0.968

Ile 0.996 0.991 0.961 0.906

Leu 0.863 0.821 0.798 -0.085

Ser 0.876 0.869 0.628 0.571

Thr 0.992 0.986 0.985 0.949

Asp 0.983 0.947 0.941 0.964

Glu 0.998 0.999 0.997 0.996

Asn 0.997 0.983 0.991 0.971

Gln 0.999 1.000 0.997 0.995

Lys 0.969 0.939 0.928 0.776

Arg 0.992 0.993 0.987 0.941

Cys 0.555 0.650 0.727 0.403

Met 0.982 0.984 0.964 0.929

Phe 1.000 0.999 0.999 0.999

Tyr 0.451 0.666 0.214 0.589

Trp 0.907 0.903 0.876 0.829

His 0.991 0.992 0.966 0.978

Pro 0.991 0.992 0.984 0.947

It should be noted that the correlation coefficients for the cases of the basis spectra with noise and without noise are comparable. Thus, the approach proposed makes it possible to estimate amino acid concentrations from noisy Raman spectra of the mixture using basis spectra with noise, and this can be done almost as efficiently as if using pure basis spectra without noise.

Nevertheless, the presence of fluorescence in the spectra is not the only problem. The form of the spectrum, that is, the presence of Raman peaks in certain wavelength ranges, has a strong influence on the result of the decomposition of the mixture spectrum. For example, in the cases of SNR = 1 and SNR = 2, one can see extremely low negative correlation values for leucine (Leu) and methionine (Met) (see Tables 4 and 5). It can be explained by the fact that the Raman spectra of these amino acids do not contain peaks that are unique to only these amino acids; moreover, the spectra are noisy. Thus, the MCR method is not efficient enough in this case.

Having studied the effect of noise in Raman spectra on the reconstruction of the concentration of amino acids in the artificially modelled mixtures, we should check the obtained relations for real mixtures, when the noise is of different nature. It is very clear that the real mixture samples (i.e., biological tissue samples) may contain

other components such as lipids and proteins. In this case, the resolved spectra of the components may have some background contribution incorporated, a final estimation of component concentrations should be performed by fixing the pure spectra of the amino acids and leaving free the background contribution, which requires the correction of the protocol settings. In addition, the efficiency of reconstruction may be affected by removing background fluorescence and normalization of the basis spectra in order to equalize their intensities. Therefore, our future research will be devoted to experiments on real mixtures of amino acids and study of the influence of basis spectra preprocessing.

Disclosures

All authors declare that there is no conflict of interests in this paper.

Acknowledgments

This work was supported by the RFBR Grant № 19-5218001 Bolg_a.

References

1. Cancer, World Health Organisation, 3 March 2021 (accessed 5 April 2021). [https://www.who.int/news-room/fact-sheets/detail/cancer].

2. A. Kubota, M. M. Meguid, and D. C. Hitch, "Amino acid profiles correlate diagnostically with organ site in three kinds of malignant tumors," Cancer 69(9), 2343-2348 (1992).

3. Y. Miyagi, M. Higashiyama, A. Gochi, M. Akaike, T. Ishikawa, T. Miura, and M. Moriyama, "Plasma free amino acid profiling of five types of cancer patients and its application for early detection," PloS One 6(9), e24143 (2011).

4. M. Shingyoji, T. Iizasa, M. Higashiyama, F. Imamura, N. Saruki, A. Imaizumi, and M. Yamakado, "The significance and robustness of a plasma free amino acid (PFAA) profile-based multiplex function for detecting lung cancer," BMC Cancer 13(1), 77 (2013).

5. Q. H. Zhao, Y. Cao, Y. Wang, C. L. Hu, A. L. Hu, L. Ruan, and M. Ren, "Plasma and tissue free amino acid profiles and their concentration correlation in patients with lung cancer," Asia Pacific journal of clinical nutrition 23(3), 429436 (2014).

6. A. M. Proenza, J. Oliver, A. Palou, and P. Roca, "Breast and lung cancer are associated with a decrease in blood cell amino acid content," The Journal of Nutritional Biochemistry 14(3), 133-138 (2003).

7. I. A. Bratchenko, L. A. Bratchenko, A. A., Moryatov, Y. A., Khristoforova, D. N. Artemyev, O. O. Myakinin, A. E. Orlov, S. V. Kozlov, and V. P. Zakharov, "In vivo diagnosis of skin cancer with a portable Raman spectroscopic device," Experimental Dermatology 30(5), 652-663 (2021).

8. A. de Juan, R. Tauler, "Multivariate Curve Resolution: 50 years addressing the mixture analysis problem - A review," Analytica Chimica Acta 1145(8), 59-78 (2020).

9. M. Garrido, F. X. Rius, and M. S. Larrechi, "Multivariate curve resolution-alternating least squares (MCR-ALS) applied to spectroscopic data from monitoring chemical reactions processes," Analytical and Bioanalytical Chemistry 390(8), 2059-2066 (2008).

10. H. Noothalapati, K. Iwasaki, and T. Yamamoto, "Biological and medical applications of multivariate curve resolution assisted Raman spectroscopy," Analytical Sciences 33(1), 15-22 (2017).

11. H. Xu, B. W. Rice, "In-vivo fluorescence imaging with a multivariate curve resolution spectral unmixing technique," Journal of Biomedical Optics 14(6), 064011 (2009).

12. P. H. Chen, R. Shimada, S. Yabumoto, H. Okajima, M. Ando, C. T. Chang, L. T. Lee, Y. K. Wong, A. Chiou, and H. O. Hamaguchi, "Automatic and objective oral cancer diagnosis by Raman spectroscopic detection of keratin with multivariate curve resolution analysis," Scientific Reports 6(1), 1-9 (2016).

13. K. Iwasaki, A. Araki, C. M. Krishna, R. Maruyama, T. Yamamoto, and H. Noothalapati, "Identification of Molecular Basis for Objective Discrimination of Breast Cancer Cells (MCF-7) from Normal Human Mammary Epithelial Cells by Raman Microspectroscopy and Multivariate Curve Resolution Analysis," International Journal of Molecular Sciences 22(2), 800 (2021).

14. J. Felten, H. Hall, J. Jaumot, R. Tauler, A. De Juan, and A. Gorzsas, "Vibrational spectroscopic image analysis of biological material using multivariate curve resolution-alternating least squares (MCR-ALS)," Nature protocols 10(2), 217-240 (2015).

15. L. A. Bratchenko, I. A. Bratchenko, A. A. Lykina, M. V. Komarova, D. N. Artemyev, O. O. Myakinin, A. A. Moryatov, I. L. Davydkin, S. V. Kozlov, and V. P. Zakharov, "Comparative study of multivariative analysis methods of blood Raman spectra classification," Journal of Raman Spectroscopy 51(2), 279-292 (2020).

16. SpectraSuite, Ocean Insight (accessed 29 July 2020). [https://oceanoptics.ru/spectrometers/59-spectrasuite.html].

17. I. A. Matveeva, O. O. Myakinin, Y. A. Khristoforova, I. A. Bratchenko, E. N. Tupikova, and V. P. Zakharov, "Possibilities for decomposing Raman spectra of amino acids mixture by Multivariate Curve Resolution (MCR) analysis," SPIE Proceedings 11582, 115821G (2020).