Научная статья на тему 'Influence of Transport Properties on Energy Resolution of Planar TlBr and CdZnTe Gamma-Ray Detectors: Monte Carlo Investigation'

Influence of Transport Properties on Energy Resolution of Planar TlBr and CdZnTe Gamma-Ray Detectors: Monte Carlo Investigation Текст научной статьи по специальности «Электротехника, электронная техника, информационные технологии»

CC BY
102
42
i Надоели баннеры? Вы всегда можете отключить рекламу.

Аннотация научной статьи по электротехнике, электронной технике, информационным технологиям, автор научной работы — Skrypnyk A.I., Khazhmuradov M.A.

The response of TlBrand CdZnTedetectors to gamma-rays was simulated by Monte Carlo method via Geant4 package. We studied the influence of transport parameters of electrons and holes on energy resolution of detectors. The modification of photopeaks with a changing the ratio of the electron and hole mobility-lifetime products was investigated. All results obtained for TlBr detectors were compared with the results for CdZnTe detectors. The efficiency for detecting gamma-quanta in the range of energies from 10 keV to 3 MeV by both kinds of detector was researched.

i Надоели баннеры? Вы всегда можете отключить рекламу.
iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.
i Надоели баннеры? Вы всегда можете отключить рекламу.

Текст научной работы на тему «Influence of Transport Properties on Energy Resolution of Planar TlBr and CdZnTe Gamma-Ray Detectors: Monte Carlo Investigation»

Influence of Transport Properties on Energy Resolution of Planar TlBr and CdZnTe Gamma-Ray Detectors: Monte Carlo Investigation

Skrypnyk A.I., Khazhmuradov M.A.

Abstact - The response of TlBr- and CdZnTe- detectors to gamma-rays was simulated by Monte Carlo method via Geant4 package. We studied the influence of transport parameters of electrons and holes on energy resolution of detectors. The modification of photopeaks with a changing the ratio of the electron and hole mobility-lifetime products was investigated. All results obtained for TlBr detectors were compared with the results for CdZnTe detectors. The efficiency for detecting gamma-quanta in the range of energies from 10 keV to 3 MeV by both kinds of detector was researched.

I. Introduction

For many years, investigation of wide band-gap semiconductors (CdZnTe, TlBr, HgI2 and other) is directed to the development of gamma-ray detectors working at room temperatures without additional cooling. However, some features of these semiconductor materials create problems in determining a detector’s main operating characteristics. Considerable non-uniformity of electrophysical characteristics of single-crystals is one of the most important factors restraining progress in achieving this goal. The most unstable characteristics include specific resistance of detector and product of mobility p and mean drift time t for electrons and holes - (pT)e,h (transport parameters of charge carriers). Planar gamma-ray detectors based on wide band-gap semiconductors have considerable spread of (pT)e,h values even if they are produced from one ingot [1]. At the same bias voltage, Ub, the characteristics of such detectors with the same sizes such as sensitivity to the registered radiation 5 and charge collection efficiency (CCE) will be different. Experiments conducted by Suzuki et al. [1] showed that, for example, in CdZnTe single-crystals, the ratio of charge transport parameters for electrons and holes within the same ingot may vary profoundly from 10 to 100.

Manuscript received November 12, 2013.

Skrypnyk A.I. is with the National Science Center Kharkov Institute of Physics and Technics. Address: Ukraine, 61108, Kharkov, Academicheskaya Str., 1, tel. (057)335-65-94 (e-mail: [email protected]).

Khazhmuradov M.A. is with the National Science Center Kharkov Institute of Physics and Technics. Address: Ukraine, 61108, Kharkov, Academicheskaya Str., 1, tel. (057)335-65-94 (email: [email protected])

Furthermore, modification of the ingot’s (pT)e,h product can be due to technological processing of the material during manufacturing into a detector, or may result from the accumulation of defects during growth or operation [2].

For a study of features of room-temperature semiconductor devices which are manufactured for detecting nuclear radiation and measuring the characteristics of the radiation fields Monte Carlo simulation can be used. A computer experiment helps to overcome the difficulties that are present as at the study of features of semiconductors as at the development of detectors based on them. In the present work, we studied the influence of the (pT)e,h products and (pT)e/(pT)h ratio on the spectroscopic characteristics of CdZnTe and TlBr- detectors using Geant4 simulation package. The detailed Monte Carlo investigation allowed us to model the response function of planar spectrometers in the gamma-ray energy range to 3 MeV. The dynamics of response function of TlBr-detectors for gamma-ray energies of 122 keV, 136 keV (57Co source) and 661.7 keV (137Cs source) was explored and compared with the dynamics of response function of CdZnTe detectors for the same gamma-ray energies. We presented how change of (pT)e/(pT)h ratio at constant value of (pT)e influences on the high and width of all simulated photopeaks. It was determined and investigated the theoretical energy resolutions of TlBr and CdZnTe detectors.

II. Description of the model

We simulated the passage of gamma-quanta through the detector by Monte Carlo method via the user program code described detail in [3], embedded in Geant4 package -universal toolkit for the simulating the passage of charged particles, neutrons and gamma-quanta through matter. The simulation procedure is divided into 2 parts. Initially, the program calculates the value of the ionization energy, Ei, transferred to the detector by the absorbed gamma-quantum with the initial energy of ET Then we calculate the value of charge induced on the detector’s contacts for every interacted photon. The computer model of the detector is approximated as much as possible to a real spectrometric device. It takes into account the statistical effects of pair generation within the detector’s volume and the modification in the amplitude of the output pulse under the influence of the electronic noise and charge-carrier capture [3].

To verify the described model we applied experimental data from 6x6x3 mm3 planar Cdo.9Zno.1Te detectors, equipped with ohmic contacts. The bias voltage, Ub, was 300 V. The electron mobility-lifetime product (px)e was selected as 3x10" 3 cm2/V. We specified the total level of noise in the CdZnTe spectrometry systems (Equivalent Noise Charge - ENC) at about 300 e- (electron charge units). The detector’s dark current was taken as 3 nA. CdZnTe detector was irradiated by 137Cs. Fig. 1 presents calculated and experimental response functions of CdZnTe detectors from 137Cs source. Overall, it is evident that used model is in good agreement with the experimental measurements.

III. Analysis of efficiency of charge collection

One of the main problems of wide band-gap semiconductor detectors results from considerable spread of (px) values for electron and holes. Transport parameters for electrons and holes directly influence on charge collection efficiency. The change in the charge collection efficiency, by turn, leads to distortions to the pulse height spectrum.

Therefore, to interpret correctly the investigated spectra it is necessary to analyze CCE. We will consider uniformity of distribution of electric field within detector. In this case, the efficiency of charge collection in the planar detector irradiated by gamma-quanta from the negative contact is described by Hecht model [4]:

П( z, pe, xe, p,, xh, d ,U) = Q- =

^ gen

(PT)e Ul

(

d2

(

(d - z) d

(Рт)е UI

(PT)hUb

d2

zd

b h W

(PT)hUb

b h

(1)

1 - exp -

V V

( (

1 - exp -

V V

Here, n is the charge-collection efficiency; Qmd is the charge induced on the detector contacts; Qgen is the average charge created at absorption energy E;, Qgen = Ej/є; d is the detector’s

thickness; and z is the depth of the gamma-quantum interaction within the detector’s material (0 < z < d).

In the following, we suppose that the values of (px)e, Ub,

and d are constant. We use the notations X

e

(PT)e Ub d

corresponding to the electron’s mean-free-path which is

(PT)h

assumed constant in this analysis, and к =

(Px)

h, 0 < к <rc .

Then, equation (1) can be rewritten in the equivalent form [3]:

П(к, z) = Xe d

1 - exp

X d - z^1

kX„

f

1 - exp

z

kX

V e j

\\

The first derivative of equation (2)

-e h

dp(K, z)

d к

(2)

is equal to

z=const

d п(к, z) = Xe - + 1 ^ 3 f-—1

d к z=const d V d Kd h V KX e h

(3)

z

The first derivative (3) is positive in the whole range,-> 0

kX e

, so that the efficiency of charge collection is a monotonically increasing function of the ratio к and respectively, (Px)h. It is correct for all z in the range from 0 to d.

In the following section, we check this statement for TlBr and CdZnTe detectors.

IV. Influence of transport parameters of TlBr- and CdZnTe- detector on its spectroscopic

CHARACTERISTICS

In the present work, we simulated 2.7x2.7x2 mm3 planar TlBr detectors equipped with ohmic contacts. Given thickness of the TlBr crystal was selected as typical for detectors based on this material. Moreover, TlBr detector with such thickness gives higher efficiency for detecting gamma-quanta with 122-keV and 136-keV energies from 57Co («97% and «93%, respectively) compared with lower thickness detector (Fig. 2).

Fig. 2. Efficiency of gamma-ray TlBr-detectors with different

thicknesses

In the range of energies of gamma-quanta less than 1 MeV, the gamma ray detector efficiency of 2-mm TlBr detector slightly exceeds this efficiency of 3-mm CdZnTe detector (Fig. 3). Therefore, as evident from Fig. 3, efficiency of detection of 661.7-keV gamma-quanta from 137Cs radioactive source does not exceed 15% for both materials with above mentioned thicknesses.

Fig. 3. Comparison of gamma ray detector efficiency for TlBr-and CdZnTe- detectors

Fig. 4 shows the transformation of 57Co spectrum obtained by simulation of TlBr detector with decreasing к value. The bias voltage, Ub, was 400 V. We specified the total level of noise in the TlBr spectrometry systems at about 400 e-. It was assumed that the detector’s dark current was 3 nA. We considered the material with values of the electron and hole mobility of 30 cm2/(Vs) and 4 cm2/(Vs), respectively, which reflect the measured gx products for electrons and holes.

Fig. 4. Transformation of 57Co spectrum obtained by TlBr-detector with decreasing к value

The electron mobility-lifetime was fixed at 5M0-4 cm2/V. The mobility-lifetime values for holes were varied from 1 ><10-4 [5] to 5M0-6 cm2/V. From Fig. 4, we observe a little shift of the centroids of 122-keV and 136-keV photopeaks in the direction of lower energy with decreasing a value of к i.e. with decreasing a value of (gx)h at const (gx)e. Moreover, the theoretical energy resolution of the studied TlBr detector for gamma-quantum energy at 122 keV declines from 5.7% to 7.6% with increasing 1/к value from 5 to 100. 136-keV photopeak tends to full degeneration. 14.4-keV photopeak

corresponded to third principal gamma-ray line of 57Co almost does not change.

Spectrum of gamma-quanta from 57Co source obtained by simulation of TlBr-detector was compared with such spectrum received by simulation of CdZnTe detector [3].

Fig. 5. Transformation of 57Co spectrum obtained by CdZnTe detector with decreasing к value

From Fig. 4 and 5, it is evident that these spectra of gamma-quanta obtained by both detectors have similar tendency to decreasing and broadening of 122-keV and 136-keV photopeaks and similar shape in this range of energies with decreasing a value of hole mobility-lifetime product. The theoretical energy resolution of CdZnTe detector for gamma-quantum energy at 122 keV drops from 1.8% to 2.3% when the value of 1/к changes from 10 to 60. The spectrum obtained by CdZnTe also has almost constant 14.4-keV photopeak (it is not shown in Fig. 5). It agrees with the experimental data of Sato et al. [Fig. 6, Ref. 6]. This reflects the fact that the depth of absorption of the main part of the gamma-quanta with 14.4-keV energy in both TlBr- and CdZnTe- materials is near to a hole-drift-length even at the worst (gx)h values.

However, irradiation of investigated TlBr detector by gamma-quanta from 57Co source gives high enough peaks in the spectrum in the range between about 30-50 keV in contrast with CdZnTe detector. The centroids of these photopeaks also are shifted in the direction of lower energy with decreasing a value of к. We suppose that they are escape peaks corresponding to gamma-radiation from ^-shells of Tl. Our simulation results agree with results of real experiment for TlB-detectors [5].

Consequently, it is evident that for gamma-quantum energies less than 150 keV, planar CdZnTe detectors of 3 mm thickness retain satisfactory spectrometric properties in the ratio range (gx)e/(gx)h below 30. Energy resolution of the investigated TlBr detectors of 2 mm thickness approximately in two times worse compared with CdZnTe detectors in this range of gamma-quanta energies. It agrees with experimental data for TlBr detectors which are obtained to the present time.

Fig. 6 presents the simulated spectrum of gamma-quanta from 137Cs source obtained for investigated TlBr-detector.

Fig. 6. Transformation of 137Cs spectrum obtained by TlBr detector with decreasing к value

In this case of investigation, because of low efficiency for detecting the gamma-quantum energy at 661.7 keV we needed to increase the number of simulated gamma-quantum trajectories of 137Cs source compared with 57Co to 108 to collect statistics. The theoretical energy resolution of the investigated TlBr-detector at a gamma-ray energy of 661.7 keV drops from 2.6% to 4.8%, when the value of 1/k changes from 5 to 20.

From Fig. 7 we observe faster degeneration of 661.7-keV photopeak in TlBr compared with 122-keV photopeak. These results agree with results of simulation of CdZnTe detector irradiated by 137Cs source. Fig. 7 shows the changes that occur around the 661.7-keV photopeak with the simulated spectrum of the 137Cs source for a CdZnTe detector [3]. The value of 1/к = 20 can be considered as the threshold level. The theoretical energy resolution of the investigated CdZnTe detector at 661.7 keV declines from 1.1% to 1.5% in the range of 1/к values from 10 to 20. The planar CdZnTe detectors with higher value of 1/к are unsuitable for the spectrometry of high-energy gamma-quanta, because even low accumulation of radiation traps can lead to the disappearance of the photopeak.

Fig. 7. Transformation of 137Cs spectrum obtained by CdZnTe detector with decreasing к value

The faster degeneration of the 661.7-keV photopeak in CdZnTe detectors compared with the 122-keV photopeak is connected with the fact that in the simulated detector the interaction of 122-keV gamma-quanta within the detector material mainly occurs in the first one-third of its thickness. Gammas with energy of 661.7 keV uniformly interact with detector throughout its entire thickness. The efficiency of charge collection, Eq. (1), depends on interaction depth. Therefore, decreasing the hole-drift-length relative to the electron-free path more strongly reduces п(к, z) and the pulse amplitude at greater depths. The full absorption cross-section of CdZnTe is small in the energy region EY above 100 keV. Therefore, this small total pulse-number from the full absorption of 661.7-keV gamma-quantum is spread therewith over a wider range of amplitudes.

It was determined that planar CdZnTe detectors theoretically can ensure an energy resolution of better than 2% at 661.7 keV provided that the value of 1/к is less than 20. In the range 1/к from 20 to 60, the detector’s resolution quickly deteriorates to 10-12% after the complete disappearance of the 661.7-keV photopeak.

From simulation of TlBr detectors, we concluded that in the range 1/k less than 10 we may receive energy resolution better than 3%. At a value of 1/k more than 45 the 661.7-keV photopeak cannot be observed by TlBr-detector.

V. Conclusion

Basing on the mentioned model, response of semiconductor TlBr-detectors to gamma-quanta from 57Co and 137Cs sources was simulated and then was compared with response of CdZnTe-detectors to gamma-quanta from the same sources. The detection efficiency of both materials was investigated. We conclude that TlBr- and CdZnTe- detectors have the same detection efficiency in the range of energies to 60 keV. At gamma-ray energies of 122 keV and 136 KeV (from 57Co source) TlBr detector has higher detection efficiency compared with CdZnTe detector. For 661.7-keV energy for both investigated detectors the given value does not exceed 15%.

We determined that investigated detectors have similar tendency to broadening of 122-keV, 136-keV and

661.7-keV photopeaks and deterioration of energy resolution with decreasing a value of (px)h at a constant value of (px)e. It is concluded that the spectroscopic properties of the both kinds of detector are maintained when the range of the (px)e/(px)h ratio is below 20. For CdZnTe if the (px)e/(px)h ratio is above 60, then the 661.7-keV photopeak cannot be observed for planar detectors, even with very low levels of electronic noise. In the case of TlBr-detectors, complete degeneration of 661.7-keV photopeak is observed at a little less (px)e/(px)h ratio. These criteria establish quality-growth requirements for spectrometric TlBr and CdZnTe materials.

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

References

[1] Suzuki M., Tashiro M., Sato G., Watanabe S., Nakazawa K., Takahashi T., Okada Y., Takahashi H., Parsons A., Barthelmy S., Cummings J., Gehrels N., Hullinger D., Krimm H., Tueller J. Hard X-Ray Response of CdZnTe Detectors in the Swift Burst Alert Telescope // IEEE Trans. Nucl. Sci. 52. 2005. P. 1033-1035.

[2] Cavallini A., Fraboni B., Dusi W., Auricchio N., Chirco P., Zanarini M., Siffert P., Fougeres P. Radiation effects on II-VI compound-based detectors // Nucl. Instr. and Meth. A 476. 2002. P. 770-778.

[3] Zakharchenko A., Rybka A., Kutny V., Skrypnyk A., KhazhmuradovM., FochukP., Bolotnikov A., James R. Transport properties and spectrometric performances of CdZnTe gamma-ray detectors // Proc. of SPIE. 2012. V. 8507, P. 85071I-1-7.

[4] Akutagawa W., Zanio K. Gamma response of semi-insulating material in the presence of trapping and detrapping // J. Appl. Phys. 40. 1969. P. 3838-3854.

[5] Shorohov M., Kouznetsov M., Lisitskiy I., Ivanov V., Gostilo V. and Owens A., Member, IEEE. Recent Results in TlBr Detector Crystals Performance // IEEE Transactions on Nuclear Science. 2009. Vol. 56, No 4, P. 1855-1858.

[6] Sato G., Parsons A., Hullinger D., Suzuki M., Takahashi T. et al. Development of a spectral model based on charge transport for the Swift/BAT 32K CdZnTe detector array // Nucl. Instr. and Meth. A 541. 2005. P. 372-384.

i Надоели баннеры? Вы всегда можете отключить рекламу.