Научная статья на тему 'DETERMINATION OF THE EFFICIENCY OF THE IKFA CORROSION INHIBITOR IN THE AGGRESSIVE ENVIRONMENT OF MDEA ABSORBENT ON A SAMPLE OF STEEL 09G2S'

DETERMINATION OF THE EFFICIENCY OF THE IKFA CORROSION INHIBITOR IN THE AGGRESSIVE ENVIRONMENT OF MDEA ABSORBENT ON A SAMPLE OF STEEL 09G2S Текст научной статьи по специальности «Технологии материалов»

CC BY
37
13
i Надоели баннеры? Вы всегда можете отключить рекламу.
Ключевые слова
MDEA (METHYLDIETHANOLAMINE) / CORROSION / UROTROPINE / HYDROCHLORIC ACID / ACETONE / DESICCANT / STEEL / DRYING CABINET / CORROSION INHIBITOR

Аннотация научной статьи по технологиям материалов, автор научной работы — Akhmedov Vohid, Rayimov Zukhriddin

In the chemical industry, the corrosion process occurring in technological devices is one of the urgent problems. Because corrosion causes great economic damage due to the rapid failure of metal products, equipment, structures and devices. Various methods and corrosion inhibitors are used to reduce this level of damage. This scientific article presents data on the effectiveness of the newly synthesized corrosion inhibitor IKFA.

i Надоели баннеры? Вы всегда можете отключить рекламу.
iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.
i Надоели баннеры? Вы всегда можете отключить рекламу.

Текст научной работы на тему «DETERMINATION OF THE EFFICIENCY OF THE IKFA CORROSION INHIBITOR IN THE AGGRESSIVE ENVIRONMENT OF MDEA ABSORBENT ON A SAMPLE OF STEEL 09G2S»

№ 5 (110)

UNIVERSUM:

ТЕХНИЧЕСКИЕ НАУКИ

май, 2023 г.

CHEMICAL ENGINEERING

DOI - 10.32743/UniTech.2023.110.5.15410

DETERMINATION OF THE EFFICIENCY OF THE IKFA CORROSION INHIBITOR IN THE AGGRESSIVE ENVIRONMENT OF MDEA ABSORBENT ON A SAMPLE OF STEEL 09G2S

Vohid Akhmedov

Professor

of the Bukhara Institute of Engineering and Technology, Republic of Uzbekistan, Bukhara E-mail: [email protected]

Zukhriddin Rayimov

Doctorant

of the Bukhara Institute of Engineering and Technology, Republic of Uzbekistan, Bukhara E-mail: [email protected]

ОПРЕДЕЛЕНИЕ ЭФФЕКТИВНОСТИ ИНГИБИТОРА КОРРОЗИИ ИКФА В АГРЕССИВНОЙ СРЕДЕ АБСОРБЕНТА МДЭА НА ОБРАЗЦЕ СТАЛИ 09Г2С

Ахмедов Вохид Низомович

профессор,

Бухарского инженерно-технологического института, Республика Узбекистан, г. Бухара

Райимов Зухриддин Хайриддин угли

докторант,

Бухарского инженерно-технологического института, Республика Узбекистан, г. Бухара

ABSTRACT

In the chemical industry, the corrosion process occurring in technological devices is one of the urgent problems. Because corrosion causes great economic damage due to the rapid failure of metal products, equipment, structures and devices. Various methods and corrosion inhibitors are used to reduce this level of damage. This scientific article presents data on the effectiveness of the newly synthesized corrosion inhibitor IKFA.

АННОТАЦИЯ

В химической промышленности процесс коррозии, происходящий в технологических устройствах, является одной из актуальных проблем. Потому что коррозия наносит большой экономический ущерб из-за быстрого выхода из строя металлических изделий, оборудования, конструкций и устройств. Для снижения этого уровня повреждений используются различные методы и ингибиторы коррозии. В данной научной статье представлены данные об эффективности вновь синтезированного ингибитора коррозии ИКФА.

Keywords: MDEA (methyldiethanolamine), corrosion, urotropine, hydrochloric acid, acetone, desiccant, steel, drying cabinet, corrosion inhibitor.

Ключевые слова: МДЭА (метилдиэтаноламин), коррозия, уротропин, соляная кислота, ацетон, эксикатор, сталь, cушильный шкаф, ингибитор коррозии.

Introduction

The damage caused by corrosion of metal products, equipment, structures and devices used in developing industries is unprecedented. For example, in the United States of America, the damage caused by corrosion is

about 300 billion dollars per year, which is 6% of the country's national income.

In the Russian Federation, 12% of the total mass of the metal stock is lost every year due to corrosion. This indicator is 30% of the annual metal production.

Библиографическое описание: Akhmedov V.N., Rayimov Z.X. DETERMINATION OF THE EFFICIENCY OF THE IKFA CORROSION INHIBITOR IN THE AGGRESSIVE ENVIRONMENT OF MDEA ABSORBENT ON A SAMPLE OF STEEL 09G2S // Universum: технические науки : электрон. научн. журн. 2023. 5(110). URL: https://7universum. com/ru/tech/archive/item/15410

№ 5 (110)

UNIVERSUM:

ТЕХНИЧЕСКИЕ НАУКИ

In oil and gas processing enterprises, heat exchange devices make up about 50% of the total devices.

30% of the total metal consumption corresponds to heat exchange devices in oil and gas processing enterprises [1].

In addition to such direct damage, there are also many indirect damages. They include the loss of power of metal equipment, damages caused by their forced shutdown due to accidents, as well as expenses for eliminating the consequences of accidents that often lead to environmental disasters.

Corrosion problems are common in shell-and-tube heat exchangers, which are used in gas absorption purification technology in gas processing plants.

These shell-and-tube heat exchangers are made of 09G2C steel alloyed with manganese and silicon.

Aggressive medium consisting of MDEA (methyl-diethanolamine) and sulfuric acid enters the shell-and-tube heat exchanger with a temperature of 50°C and leaves

май, 2023 г.

the device with a temperature of 80oC. The pressure in the heat exchanger is 1.66 kg s/cm2 [2].

The proposed IKFA corrosion inhibitor is directly added to MDEA at a mass ratio of 0.05%. This composition blocks the corrosion process that occurs in the heat exchanger during the circulation of natural gas in the process of absorption purification [3].

Below are the results of determining the corrosion rate of 09G2C steel samples and the effectiveness of IKFA corrosion inhibitor based on GOST 9.905-82.

Research methodology

First experiment: Determining the corrosion rate of 09G2C steel samples in the aggressive environment of MDEA without an inhibitor.

To perform the test, the following sequence of operations was performed:

l.Before starting the test, the mass of 3 specially numbered steel samples was determined (Fig. 1 A). The obtained results were recorded in the table. (Table 1)

(A) a

Figure 1. 09G2C steel samples (A) and gauge dimensions (B) Pre-test masses of steel samples

Table 1.

Data

Samples Masses,m1 (g)

First sample 58,48

Second sample 67,60

Third sample 75,80

2. The overall dimensions of each sample are determined (Fig. 1 B). The obtained results were recorded in the table. (Table 2)

Table 2.

Dimensions of steel samples

First sample (mm) Second sample Third sample

a 40 41 42

b 18 20 21

c 11 11 11

№ 5 (110)

UNIVERSUM:

ТЕХНИЧЕСКИЕ НАУКИ

май, 2023 г.

3. Their surfaces were calculated based on the overall dimensions of the samples (1).

2(a*b + b*c + a*c)

(1)

First sample: 2(40 x 18+18 x 11+40 x 11 )=2716mm2 Second sample : 2(41 x20+20 x 11+41 x 11 )=2982mm2

Third sample: 2(42 x21+21 x 11+42 x 11 )=3150mm2

4. The steel samples were polished until there were no sharp edges on the sandpaper.

5. Polished steel samples were kept for 4 minutes in a container containing "cleaning solution" (urotropin 10 g + hydrochloric acid 100 g) (Fig. 2). The temperature of the cleaning solution is 15-25° [4].

Figure 2. Cleaning solution

6. Steel samples were scraped with a eraser a stream of distilled water.

7. Then the surface of the steel samples was wiped with acetone and wrapped in filter paper.

8. After that, it was placed in a drying cabinet at a temperature of 100° for 2 hours.

9. It was then kept in a desiccator until it cooled.

10. After that, the steel samples were placed in a 500 ml conical flask with a temperature of 75 °C,

300 ml of MDEA saturated with absorbent, and a reflux condenser (Fig. 3). [5]

11. The time was recorded. The duration of the test was 24 hours.

12. Steel samples were taken after 24 hours and cooled.

13. Then, the mass of each steel sample was determined and the results were recorded in the table. (Table 3)

Table 3.

Masses of steel samples after testing

Samples Masses,m2 (g)

First sample 58,414

Second sample 67,53

Third sample 75,73

14. Based on the obtained results, the corrosion rate was calculated (2).

ml — ml

Karav =

SXT

(2)

Here: m1 = mass of the steel sample before the test; m2 = mass of the steel sample after the test; S = steel sample surface, cm2; T=test time, hours.

First sample: Kgrav = 58,48 58,414 = 10,1*10-5g/sm2

Second sample: Kgrav

67,60-67,53 29.82x24

= 9,781*10-5g/sm2

Third sample: Kgrav = 75,8" 75,73 = 9,259*10"5g/sm:

iНе можете найти то, что вам нужно? Попробуйте сервис подбора литературы.

The average result of three samples was equal to 9,7*10-5 g/sm2. [6] Therefore, the corrosion rate of 09G2C steel samples in the aggressive environment of MDEA without inhibitor Kgrav=9,7*10-5 g/sm2.

№ 5 (110)

UNIVERSUM:

ТЕХНИЧЕСКИЕ НАУКИ

май, 2023 г.

1-steel samples; 2-conical flask; 3-return cooler; 4-heater; 5-tripod Figure-3. Test device

Second experiment: Determination of corrosion rate of 09G2C steel samples in an aggressive environment with IKFA corrosion inhibitor of MDEA.

To carry out the test, the same sequence of work as in the first experiment is performed. The testing process lasted 24 hours.

The pre-test and post-test masses of steel samples were determined in the appropriate order and the results were recorded (Table 4).

Table 4.

Masses of steel samples

Before the test (gr) After the test (gr)

First sample

mi=61,02 m2=61,01

Second sample

mi=67,69 m2=67,68

Third sample

mi=70,21 m2=70,21

№ 5 (110)

UNIVERSUM:

ТЕХНИЧЕСКИЕ НАУКИ

май, 2023 г.

The results of determining the dimensions of each sample were recorded in the table below. (Table 5)

Table 5.

Dimensions of steel samples

First sample (mm) Second sample (mm) Third sample (mm)

a 37 40 40

b 19 20 21

c 10 11 10

Their surfaces were calculated based on the overall dimensions of the samples (1).

First sample: 2(37x19+19x10+37x10)=2526mm2 Second sample: 2(40x20+20x11+40x11)=2920mm2 Third sample: 2(40 x21+21 x 10+40 x 10)=2900mm2

Based on the obtained results, the corrosion rate was calculated (2).

First sample: Kgrav =

61,02-61,01

1,65*10-5g/sm2

25,26x24

Second sample: Kgrav = ^ ^^ = 1,42*10-5g/sm2

Third sample: Kgrav = 70,21 70,21 = o g/sm2

r & 29,00x24 &

The average result of three samples was equal to 1.02 g/cm2. So, the corrosion rate of 09G2C steel samples in an aggressive environment with IKFA corrosion inhibitor of MDEA is Kgrav=1.02 g/cm2.

Summary

The results of the analysis showed that the corrosion rate of the tested steel samples in the aggressive environment of MDEA without added corrosion inhibitor was equal to 8g/cm2.

In the second experiment, the corrosion rate of steel samples tested in the aggressive environment of MDEA with IKFA corrosion inhibitor was equal to 1.02g/cm2 [7].

Therefore, when IKFA corrosion inhibitor was used, the rate of corrosion was dramatically slowed down [8].

We determine the effectiveness of IKFA corrosion inhibitor using the following formula (3):

Z = vo-ving x юр

Vo

(3)

Here: Vo = lost mass of the steel sample tested in an environment without an inhibitor; Ving=mass loss of a steel specimen tested in an inhibitory environment.

The average mass loss of a steel sample tested in an environment without an inhibitor: Vo=0.0566;

The average mass loss of a steel sample tested in an inhibitory environment: Ving=0.0066;

Z =

0,0687-0,0066 0,0687

x 100=90.4%

Therefore, the effectiveness of IKFA corrosion inhibitor in the aggressive environment of sulphide acid of MDEA was 90.4%.[9]

References:

1. Ахмедов Вохид Низомович, Бобир Баходир Угли Олимов, and Шомурод Комилович Назаров. "Электронная структура и квантово-химические расчёты виниловых эфиров фенолов." Universum: химия и биология 4 (70) (2020).

2. Жумаев Ж.Х., Ахмедов В., Шарипова Н.У. ВЛИЯНИЕ ПРИРОДЫ И КОЛИЧЕСТВА КАТАЛИЗАТОРА ПРИ СИНТЕЗЕ МОРФОЛИНОВЫХ НЕНАСЫЩЕННЫХ ПРОДУКТОВ ПРИ УЧАСТИИ ВИНИЛАЦЕТИЛЕНА // Москва. - 2021. - С. 58-61..

3. Zuhriddin R., & Niginabonu J. (2022). PRODUCTION OF POLYETHYLENE TEREPHTHALATE. Universum: технические науки, (5-11 (98)), 58-62.

4. Садирова С.Н., Темирова М.И., & Алиева Н.И. (2020). Исследование проквашенности каракуля с применением вторичных продуктов молочного производства. International Journal of Advanced Technology and Natural Sciences, 1(1), 39-44.

5. O'G'Li, R. Z. K., & Qizi J.N. Q. (2022). ANALYSIS OF IMPORTANCE AND METHODS OF PRODUCTION OF BLOCK SOPOLYMERS BASED ON POLYETYLENTEREPHTALATE. International Journal of Advanced Technology and Natural Sciences, 3(1), 51-55.

6. Рахматов М.С. Влияние катализатора, температуры и растворителя на синтез и выход продукта реакции с виниловым эфиром салициловый кислоты в присутствии винилацетилена //Universum: химия и биология. -2020. - №. 11-2 (77). - С. 16-20

7. Ниёзова, Раъно Нажмиддиновна. "Экологические и эксплуатационные свойства жированных кож на основе синтетических жирных кислот." Science and Education 2.12 (2021): 347-352.

8. Zuhriddin, R., Niginabonu, J., Aminjon, V., & Temurbek, D. (2022). MECHANISMS OF ETERIFICATION OF TEREFTALIC ACID WITH ETYLENGLYCOL. Universum: технические науки, (5-11 (98)), 63-67.

9. V.N.Axmedov, Z.X.Rayimov, G.A.G'afurova.Tereftal kislota hosilalari. ISBN 978-9943-9265-3-0 Monografiya "Durdona" nashriyoti. 2023. 152 bet.

i Надоели баннеры? Вы всегда можете отключить рекламу.