HIRURGIA POZVONOCHNIKA 2020;17(l):54-60 © A.V. KOSULIN ET al., 2020
application of three-level navigation template in surgery for hemivertebrae in adolescents
A.V. Kosulin, D.V. Elyakin, L.A. Kornievskiy, A.M. Darkovskaya, I.A. Bulatova, A.A. Pashko
Saint-Petersburg State Pediatric Medical University, St. Petersburg, Russia
Objective. To assess efficacy and safety of using a three-level navigation template in the surgical treatment of congenital spinal deformities. Material and Methods. Three-level navigation templates were used in surgical treatment of four consecutively operated 10—17-year-old patients with congenital scoliosis associated with thoracic hemivertebrae. The correctness of screw position was evaluated according to CT data using a 2-mm increment method: class 0 — intraosseous screw position; class 1 — the screw extends beyond the pedicle cortex by less than 2 mm; class 2 — by 2—4 mm; and class 3 — by more than 4 mm. Preoperative DICOM data were processed with free software. The model of target zone and navigation template were 3D printed and used in surgery.
Results. Four of 16 (25 %) pedicles were narrower than 4.35 mm and were estimated as difficult for implantation with a planned violation of the integrity of the endplate. Perforation of the outer cortical layer took place in all these cases, and screw position corresponded to class 2 only in pedicle width of 1.9 mm. In pedicles wider than 4.35 mm, 11 of 12 (91.7 %) screws were implanted intraosseously. One screw extended beyond the pedicle cortex by 0.8 mm (class 1).
Conclusion. Three-level navigation template can be considered as an effective means of positioning transpedicular screws in secondarily changed segments adjacent to anomalous one and confounding implantation. Free software is sufficient for preparing 3D-model of target zone and navigation template, and such a model is a highly informative reference object that is convenient to use during the operation. A navigation template made using 3D printing does not require the use of expensive equipment, which can make surgery for congenital scoliosis more accessible.
Key Words: 3D printing, navigation template, congenital scoliosis.
Please cite this paper as: Kosulin AV, Elyakin DV, Kornievskiy LA, Darkovskaya AM, Bulatova IA, Pashko AA. Application of three-level navigation template in surgery for hemivertebrae in adolescents. Hir. Pozvonoc. 2020;17(1):54—60. In Russian. DOI: http://dx.doi.org/10.14531/ss20201.54-60.
As the equipment and expendable supplies for 3D printing become less expensive and more available, this technology has been used in medicine to manufacture physical objects based on digital data more and more frequently. In surgery of the spine, cerebral and facial skull, middle ear, larynx, and heart [1-7], the object of intervention is a complex three-dimensional structure characterized by individual variability. Reproducing this three-dimensional structure is a novel imaging modality that supplements the conventional 3D imaging, simplifies preoperative planning, improves safety of the surgery, and reduces its traumaticity [8]. If high-precision implantation is needed, 3D printing will make it possible to manufacture a personified navigational tool that allows one to reproduce the directions and
distances determined during surgery simulation [9].
The difficulties related to corrective surgeries performed for patients with congenital vertebral anomalies are usually caused by the anatomical features of the anomalous segment and the adjacent ones. Whereas the general principles of hemivertebra excision and instrumented correction of the deformity are well-known [10, 11], the procedure used for transpedicular screw implantation near the deformity apex can be challenging because of the anatomical features of pedicles (their small size). The involvement of these structures increases the risk of improper screw placement, including the risk of damaging the neural structures.
The objective of this study was to assess the effectiveness and safety of application of the three-level naviga-
54
tional template in surgical treatment of congenital spine deformities.
Material and Methods
Three-level navigational templates were used in surgical treatment of four consecutively operated 10-17-year-old patients with congenital spine deformities caused by thoracic hemivertebrae.
The demographic, clinical, and radiographic data of patients enrolled in the study are shown in Table 1.
If any of the frontal pedicle dimension (height or width) during preop-erative planning was found to be less than 4.35 mm (the minimal diameter of a transpedicular screw in the standard implantation kit), screw implantation was regarded as potentially difficult (improper). The DICOM data obtained
for preoperative CT scans were processed using open-source software. A virtual 3D model of the zone of interest covering the anomalous vertebra and the adjacent ones was built using the Threshold Effect of density in the Slicer program. Primary processing of the model involving the elimination of excessive elements and spatial orientation was then conducted using the Autodesk Meshmixer program. Blender 3D editor was used to visualize the planar cross sections of the vertebrae to be implanted and perform positioning of cylinder-shaped figures (a virtual drill) along the optimal trans-pedicular trajectories. The same cylinder simulated the pedicle of a hemivertebra (Fig. 1a). In case of prognostically difficult implantation, it was planned that the drill will move juxtapedicularly and perforate the outer cortical layer. Cylindrical defects corresponding to positions of the optimal transpedicular trajectories were simulated by subtracting the figures of the virtual drill from the model. Hence, five trajectories running through the pedicle of the hemivertebra and two adjacent vertebrae were formed in each model. The supporting plate of the navigational template corresponding to three adjacent vertebrae was formed using 3D graphics primitives. The supporting plate was shaped as an imprint from the posterior vertebral structures using the projection algorithm (Shrinkwrap modifier). Hollow cylindrical figures of guiding tubes simulated the position of a virtual drill. The navigational template was created by combining them with the supporting plate (Fig. 1b). The model of the zone of interest and the navigational template were exported to the STL format and 3D printed using PLA material (Fig. 1c, d).
The ready-to-use templates were subjected to low-temperature sterilization and utilized during surgical interventions. The posterior vertebral structures were identified according to osseous landmarks using the 3D model as a reference object and thoroughly skeletonized. The supporting plate of the navigational template was then placed onto the posterior structures, and its stability was controlled. The transpedicular trajectories through
the guiding tubes were formed using the drill (Fig. 1e). The integrity of bone walls was controlled with a probe. Transpedic-ular screws (minimal diameter, 4.35 mm) were implanted into the vertebrae adjacent to the anomalous one. The body of the hemivertebra was resected by trans-pedicular eggshell osteotomy, and instrumented correction of the deformity was subsequently performed [11].
All the patients underwent postoperative CT scanning. The screw placement accuracy was assessed using the two-millimeter increment system [12]: class
0 - intraosseous screw positioning; class
1 - the screw extends beyond the pedicle cortex by less than 2 mm; class 2 - the screw extends beyond the pedicle cortex by less than 2-4 mm; and class 3 - the screw extends beyond the pedicle cortex by more than 4 mm.
Results
Table 2 shows the dimensions of the apical arch vertebrae (to be operated on) determined during preoperative planning, as well as postoperative assessment of positioning of transpedicular screws.
Hence, four (25 %) of 16 pedicles with pedicle width less than 4.35 mm were regarded as prognostically difficult for implantation (as the endplate integrity would probably be violated). Perforation of the outer cortical layer took place in all these cases (Fig. 2); class 2 screw position was observed only for pedicle width of 1.9 mm.
When the pedicle width was more than 4.35 mm, intraosseous screw positioning (class 0) was achieved in 11 (91.7 %) of 12 cases; in one case, the
screw extended beyond the cortical layer by 0.8 mm (class 1).
Discussion
Among the vast number of publications addressed to application of navigational templates in spine surgery, most studies focus on screw placement into the cervical vertebrae in adult patients [13-19]. Closely related to our study, the publication by D.N. Kokushin et al. [20] is limited to in vitro application of navigational templates to manage hemivertebrae, while we found no data on clinical application of this method in a homogeneous patient population with spinal malformations.
Most authors prefer to print both the navigational template and the model of the zone of interest [13-15, 17-19, 21-25], although it is not always regarded as an obligatory stage [16, 26-28]. In our opinion, the model of the zone of interest improves the quality of preoper-ative planning, makes it easier to explain the essence of the pathology and the upcoming surgery to patients and their representatives. Furthermore, it makes orientation in the surgical wound easier and can be used for educational purposes. These properties of the model largely depend on subjective impressions. A survey study needs to be conducted to make them unbiased.
A reported drawback of the navigational template is that the posterior vertebral structures need to be meticulously skeletonized, which causes significant blood loss and tissue injury [29]. Meanwhile, planning joint fusion surgery per se requires skeletonization of the pos-
55
spine deformities
Fig. 1
Stages of preoperative modeling and perioperative application of a three-level template: a - virtual modeling of the surgery; b - a virtual model of navigational template with five guiding tubes; c, d - the ready-to-use three-level template, dorsal (c) and ventral (d) views; e - perioperative application of the template
terior structures as one of the prerequisites for the formation of complete local bone block.
Single-level navigational templates were used in most studies: they prevent the inaccuracy related to displacement of the adjacent vertebrae as the patient changes his or her body position [13, 14, 16-19, 23, 25]. Meanwhile, R.A. Kova-lenko et al. [30] demonstrated that higher total instrumentation area increases template stability and, therefore, ensures better implantation accuracy. In particular, craniocaudal stability is improved
by increasing the distance between two marginal supporting points in the sagittal plane. Hence, when performing trans-pedicular fixation of a limited-mobility zone, it is preferred to use one navigational template, which allows more reliable positioning of the supporting plate and minimizing the time needed to consecutively insert several templates. In turn, the application of a multi-level template consisting of several supporting plates linked by flexible inserts is superfluous [29].
To prevent displacement of the adjacent vertebrae, which is an error source during navigation, it was suggested that patients undergo preoperative CT scanning in prone position; however, it is impossible to ensure that the body position inside an MRI machine precisely coincides with that on the operating table [27, 28]. Meanwhile, in patients with congenital thoracic scolio-sis caused by hemivertebrae, the apical curve formed by the hemivertebra and the adjacent vertebrae usually has limited mobility, which ensures stability of the three-level template.
The open-source software is sufficient for preparing the 3D model of the zone of interest and creating the navigational template. This model is a highly informative reference object that is convenient to use during surgery. The 3D printed navigational template does not require expensive equipment. It can be regarded as an alternative to the currently used CT navigation and allows one to conduct the type of surgeries discussed in this study at medical settings where the CT navigation equipment is not available.
Conclusions
The three-level navigational template is an effective and safe means of positioning transpedicular screws in children with thoracic hemivertebrae, including the pronounced changes in the segments adjacent to the abnormal one.
The limitations of this study include small sample size and no reference group.
All the authors have equally contributed to preparing this manuscript. The study had no sponsors. The authors declare that there is no conflict of interest.
56
Table 2 The minimal width of pedicles of the vertebrae adjacent to the anomalous one and postoperative evaluation of screw position accuracy
Patient Level and side Pedicle width, mm Class of screw placement accuracy
1 T10, left 2.7 1
T10, right 5.2 0
T12, left 6.8 0
T12, right 2.4 1
2 T2, left 5.3 0
T2, right 4.9 1
T4, left 8.3 0
T4, right 1.9 2
3 T8, left 4.9 0
T8, right 5.7 0
T10, left 4.9 0
T10, right 4.9 0
4 T8, left 2.9 1
T8, right 4.6 0
T10, left 6.7 0
T10, right 4.9 0
s d
T12
T12
Fig. 2
The vertebra adjacent to the abnormal one: preoperative (a) and postoperative (b) MR scans: a - prognostically difficult implantation on the right side (pedicle width, 2.4 mm); b - the screw perforates the outer cortical plate on the right side (class 1)
b
a
57
References
1. Wilcox B, Mobbs RJ, Wu AM, Phan K. Systematic review of 3D printing in spinal surgery: the current state of play. J Spine Surg. 2017;3:433-443. DOI: 10.21037/ jss.2017.09.01.
2. Aydin HE, Kaya I, Aydin N, Kizmazoglu C, Karakoc F, Yurt H, Husemog-
lu RB. Importance of three-dimensional modeling in cranioplasty. J Craniofac Surg. 2019;30:713-715. DOI: 10.1097/SCS.0000000000005121.
3. Filatova OO, Klimov AG, Seleznev BV. The usage of combination of tricalcium phosphate and polylactic acid as materials for 3D printing of alloplastic blocks. Pediatrician (St Petersburg). 2017;8(3):47-50. In Russian]. DOI: 10.17816/PED8347-50.
4. Vlasova GV, Pavlov PV. Congenital middle ear cholesteatoma in children: retrospective analysis of 23 cases. Vestnik otorinolaringologii. 2017;82(S5):59-60. In Russian
5. Zaharova ML, Pavlov PV. Congenital larynx diseases in children. Russian otorhino-laryngology. 2017;86(1):31-35. In Russian
6. Hamdan AL, Haddad G, Haydar A, Hamade R. The 3D printing of the paralyzed vocal fold: added value in injection laryngoplasty. J Voice. 2018;32:499-501. DOI: 10.1016/j.jvoice.2017.07.011.
7. Giannopoulos AA, Mitsouras D, Yoo SJ, Liu PP, Chatzizisis YS, Rybicki FJ. Applications of 3D printing in cardiovascular diseases. Nat Rev Cardiol. 2016;13:701-718. DOI: 10.1038/nrcardio.2016.170.
8. Tack P, Victor J, Gemmel P, Annemans L. 3D-printing techniques in a medical setting: a systematic literature review. Biomed Eng Online. 2016;15:115. DOI: 10.1186/ s12938-016-0236-4.
9. Wu AM, Lin JL, Kwan KYH, Wang XY, Zhao J. 3D-printing techniques in spine surgery: the future prospects and current challenges. Expert Rev Med Devices. 2018;15:399-401. DOI: 10.1080/17434440.2018.1483234.
10. Ulrich EV, Mushkin AYu. Surgical Correction of Congenital Spinal Abnormalities in Children. St. Petersburg, 2007. In Russian
11. Ratliff JK, Nasser R Hemivertebrae resection. In: Operative Techniques: Spine Surgery, 3rd ed. Ed. by Vaccar AR, Baron EM. Elsiever; 2018:398-402.
12. Aoude AA, Fortin M, Figueiredo R, Jarzem P, Ouellet J, Weber MH. Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review. Eur Spine J. 2015;24:990-1004. DOI: 10.1007/s00586-015-3853-x.
13. Burtsev AV, Pavlova OM, Ryabykh SO, Gubin AV. Computer 3D-modeling of patient-specific navigational template for cervical screw insertion. Hir. Pozvonoc. 2018;15(2):33-38. In Russian]. DOI: 10.14531/ss2018.2.33-38.
14. Kovalenko RA, Rudenko VV, Kashin VA, Cherebillo VYu, Ptashnikov DA. Application of patient-specific 3D navigation templates for pedicle screw fixation of subaxial and upper thoracic vertebrae. Hir. Pozvonoc. 2019;16(2):35-41. In Russian]. DOI: 10.14531/ss2019.2.35-41.
15. Deng T, Jiang M, Lei Q, Cai L, Chen L. The accuracy and the safety of individualized 3D printing screws insertion templates for cervical screw insertion. Comput Assist Surg (Abingdon). 2016;21:143-149. DOI: 10.1080/24699322.2016.1236146.
16. Kaneyama S, Sugawara T, Sumi M. Safe and accurate midcervical pedicle screw insertion procedure with the patient-specific screw guide template system. Spine. 2015;40:E341-E348. DOI: 10.1097/BRS.0000000000000772.
17. Pu X, Luo C, Lu T, Yao S, Chen Q. Clinical application of atlantoaxial pedicle screw placement assisted by a modified 3d-printed navigation template. Clinics (Sao Paulo). 2018;73:e259. DOI: 10.6061/clinics/2018/e259.
18. Sugawara T, Higashiyama N, Kaneyama S, Sumi M. Accurate and simple screw insertion procedure with patient-specific screw guide templates for posterior C1-C2 fixation. Spine. 2017;42:E340-E346. DOI: 10.1097/BRS.0000000000001807.
19. Wang F, Li CH, Liu ZB, Hua ZJ, He YJ, Liu J, Liu YX, Dang XQ. The effectiveness and safety of 3-dimensional printed composite guide plate for atlantoaxial pedicle
screw: A retrospective study. Medicine (Baltimore). 2019;98:e13769. DOI: 10.1097/ MD.0000000000013769.
20. Kokushin DN, Vissarionov SV, Baindurashvili AG, Ovechkina AV, Poznov-
ich MS. Comparative analysis of pedicle screw placement in children with congenital scoliosis: freehand technique (in vivo) and guide templates (in vitro). Travmatologiya I Ortopediya Rossii [Traumatology and Orthopedics of Russia]. 2018;24(4):53-63. In Russian]. DOI: 10.21823/2311-2905-2018-24-4-53-63.
21. Kosulin AV, Elyakin DV, Lebedeva KD, Sukhomlinova AE, Kozlova EA, Orek-hova AE. Navigation template for vertebral pedicle passage in transpedicular screw fixation. Pediatrician (St. Petersburg). 2019;10(3):45-50. In Russian]. DOI: 10.17816/ PED10345-50.
22. Cecchinato R, Berjano P, Zerbi A, Damilano M, Redaelli A, Lamartina C. Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial. Eur Spine J. 2019;28:1712-1723. DOI: 10.1007/ s00586-019-05978-3.
23. Garg B, Gupta M, Singh M, Kalyanasundaram D. Outcome and safety analysis of 3D-printed patient-specific pedicle screw jigs for complex spinal deformities: a comparative study. Spine J. 2019;19:56-64. DOI: 10.1016/j.spinee.2018.05.001.
24. Li X, Zhang Y, Zhang Q, Zhao C, Liu K. Clinical application of a drill guide template for pedicle screw placement in severe scoliosis. Acta Ortop Bras. 2017;25:67-70. DOI: 10.1590/1413-785220172502138828.
25. Lu T, Liu C, Dong J, Lu M, Li H, He X. Cervical screw placement using rapid prototyping drill templates for navigation: a literature review. Int J Comput Assist Radiol Surg. 2016;11:2231-2240. DOI: 10.1007/s11548-016-1414-3.
26. Fan Y, Du J, Zhang J, Liu S, Xue X, Huang Y, Zhang J, Hao D. Comparison of accuracy of pedicle screw insertion among 4 guided technologies in spine surgery. Med Sci Monit. 2017;23:5960-5968. DOI: 10.12659/MSM.905713.
27. Liu K, Zhang Q, Li X, Zhao C, Quan X, Zhao R, Chen Z, Li Y. Preliminary application of a multi-level 3D printing drill guide template for pedicle screw placement in severe and rigid scoliosis. Eur Spine J. 2016;26:1684-1689. DOI: 10.1007/ s00586-016-4926-1.
28. Merc M, Drstvensek I, Vogrin M, Brajlih T, Friedrich T, Recnik G. Error rate of multi-level rapid prototyping trajectories for pedicle screw placement in lumbar and sacral spine. Chin J Traumatol. 2014;17:261-266.
29. Azimifar F, Hassani K, Saveh AH, Ghomsheh FT. A medium invasiveness multilevel patient's specific template for pedicle screw placement in the scoliosis surgery. Biomed Eng Online. 2017;16:130. DOI: 10.1186/s12938-017-0421-0.
30. Kovalenko RA, Kashin VA, Cherebillo VYu, Sharifov RM, Mironchuk RR, Akopov AL, Ivanov VA. Determination of optimal design of navigation templates for transpedicular implantation in the cervical and thoracic spine: results of cadaveric studies. Hir. Pozvonoc. 2019;16(4):77-83. In Russian]. DOI: 10.14531/ss2019.4.77-83.
Address correspondence to:
Kosulin Artem Vladimirovich, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Peterburg, 194100, Russia, [email protected]
Received 13.01.2020
Review completed24.01.2020
Passed for printing 30.01.2020
58
Artem Vladimirovich Kosulin, assistant Professor, Department of Operative Surgery and Topographic Anatomy n. a. F.I. Valker, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-9505-222X, [email protected];
Dmitry Viktorovich Elyakin, pediatric surgeon, Surgical Department No.2, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-6575-7464, [email protected];
Leonid Aleksandrovich Kornievskiy, assistant professor, Department of Otorhinolaryngology, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-8635-1666, [email protected];
Aleksandra Maksimovna Darkovskaya, student, Pediatric Faculty, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-2511-9679, [email protected];
Irina Anatolyevna Bulatova, MD, PhD, associate professor, Department of Operative Surgery and Topographic Anatomy n.a. F.I. Valker, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-1112-6966, [email protected];
Anna Anatolyevna Pashko, MD, PhD, associate professor, Department of Operative Surgery and Topographic Anatomy n. a. F.I. Valker, Saint-Petersburg State Pediatric Medical University, 2 Litovskaya str., St. Petersburg, 194100, ORCID: 0000-0002-6413-6113, [email protected].
59
_HIRURGIA POZVONOCHNIKA 2020;17(l):54-60_
av. kosulin et al. application of three-level navigation template in surgery for hemivertebrae in adolescents
60
spine deformities